Evaluation of the minimum volume of salvage blood required for the successful use of two different autotransfusion devices


Baumann, Cyrille; Lamesic, Goran; Weiss, Markus; Cushing, Melissa M; Haas, Thorsten (2015). Evaluation of the minimum volume of salvage blood required for the successful use of two different autotransfusion devices. Paediatric Anaesthesia, 25(3):258-264.

Abstract

BACKGROUND Autotransfusion in pediatric anesthesia is beneficial in several clinical settings; however, more frequent usage is deterred by process-related costs and the fact that the absolute volume of blood returned may be minimal. OBJECTIVE Two autotransfusion devices (CATS, Fresenius Kabi AG, and Sorin Xtra, Sorin GmbH) with different technologies were evaluated to determine the minimum volume necessary to efficiently process salvage blood. METHODS Banked blood was diluted to simulate different clinical scenarios. Volume and red cell mass in returned blood were analyzed as a function of initial volume and red cell mass in the collection reservoir using incremental increases in the volume of blood added to the reservoir. RESULTS The volume of the returned processed blood for infusion was dependent on degree of dilution and blood volume in the reservoir. The CATS required a smaller initial minimum volume to process and successfully return; however, the Xtra provided a more efficient recovery at low initial collected volumes. Using test blood of various degrees of dilution, the CATS and Xtra were able to return blood for retransfusion after a mean initial red cell mass of 48.7 ml (±4.9 ml) and 53.7 ml (±5.0 ml) were processed, respectively. CONCLUSIONS Red cell mass rather than salvaged blood volume represents a reliable predictor of the successful use of an autotransfusion device. Measuring the hematocrit of the salvaged blood could improve the use of the devices. The investigated devices are likely to be roughly equivalent in effectiveness.

Abstract

BACKGROUND Autotransfusion in pediatric anesthesia is beneficial in several clinical settings; however, more frequent usage is deterred by process-related costs and the fact that the absolute volume of blood returned may be minimal. OBJECTIVE Two autotransfusion devices (CATS, Fresenius Kabi AG, and Sorin Xtra, Sorin GmbH) with different technologies were evaluated to determine the minimum volume necessary to efficiently process salvage blood. METHODS Banked blood was diluted to simulate different clinical scenarios. Volume and red cell mass in returned blood were analyzed as a function of initial volume and red cell mass in the collection reservoir using incremental increases in the volume of blood added to the reservoir. RESULTS The volume of the returned processed blood for infusion was dependent on degree of dilution and blood volume in the reservoir. The CATS required a smaller initial minimum volume to process and successfully return; however, the Xtra provided a more efficient recovery at low initial collected volumes. Using test blood of various degrees of dilution, the CATS and Xtra were able to return blood for retransfusion after a mean initial red cell mass of 48.7 ml (±4.9 ml) and 53.7 ml (±5.0 ml) were processed, respectively. CONCLUSIONS Red cell mass rather than salvaged blood volume represents a reliable predictor of the successful use of an autotransfusion device. Measuring the hematocrit of the salvaged blood could improve the use of the devices. The investigated devices are likely to be roughly equivalent in effectiveness.

Citations

2 citations in Web of Science®
2 citations in Scopus®
Google Scholar™

Altmetrics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Children's Hospital Zurich > Clinic for Surgery
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:2015
Deposited On:28 Nov 2014 12:10
Last Modified:05 Apr 2016 18:33
Publisher:Wiley-Blackwell Publishing, Inc.
ISSN:1155-5645
Publisher DOI:https://doi.org/10.1111/pan.12535
PubMed ID:25267461

Download

Full text not available from this repository.
View at publisher

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.