Abstract

Knowledge of the consumed amount of volatile anesthetic (VA) expressed in liquid agent is necessary to enable agent sparing dosing measures and for billing purposes. The widespread Draeger Primus™ anesthesia machine displays in its logbook the amount of consumed VA at the end of each anesthesia, but the reliability of this parameter is yet unknown. The objective was to evaluate the precision and reliability of the inbuilt VA consumption display in Draeger Primus™ anesthesia machines as compared with the gold standard of weighing the vaporizer before and after anesthesia. In this prospective laboratory investigation we compared the VA consumption displayed by the Draeger Primus™ anesthesia machine with measured vaporizer weight differences before and after 10 sevoflurane and 10 desflurane anesthesias. We assessed the average difference and spread of values between the predicted (displayed) and measured (control) values for VA consumption. The displayed sevoflurane consumption overestimated the measured values by 4.3 ± 5.4 ml (7.6 %). The displayed desflurane consumption underestimated the measured values by -3.5 ± 6.3 ml (6.2 %). Nine from 10 sevoflurane pairs of values and all desflurane pairs of values were within ±1.96 SD. The displayed VA consumption calculations for sevoflurane and desflurane in the Draeger Primus™ are sufficiently reliable to estimate the pharmacoeconomic impact of VA delivery during inhalational anesthesia.

Knowledge of the consumed amount of volatile anesthetic (VA) expressed in liquid agent is necessary to enable agent sparing dosing measures and for billing purposes. The widespread Draeger Primus™ anesthesia machine displays in its logbook the amount of consumed VA at the end of each anesthesia, but the reliability of this parameter is yet unknown. The objective was to evaluate the precision and reliability of the inbuilt VA consumption display in Draeger Primus™ anesthesia machines as compared with the gold standard of weighing the vaporizer before and after anesthesia. In this prospective laboratory investigation we compared the VA consumption displayed by the Draeger Primus™ anesthesia machine with measured vaporizer weight differences before and after 10 sevoflurane and 10 desflurane anesthesias. We assessed the average difference and spread of values between the predicted (displayed) and measured (control) values for VA consumption. The displayed sevoflurane consumption overestimated the measured values by 4.3 ± 5.4 ml (7.6 %). The displayed desflurane consumption underestimated the measured values by -3.5 ± 6.3 ml (6.2 %). Nine from 10 sevoflurane pairs of values and all desflurane pairs of values were within ±1.96 SD. The displayed VA consumption calculations for sevoflurane and desflurane in the Draeger Primus™ are sufficiently reliable to estimate the pharmacoeconomic impact of VA delivery during inhalational anesthesia.

Citations

1 citation in Web of Science®
2 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

15 downloads since deposited on 30 Dec 2014
14 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Institute of Anesthesiology
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:2015
Deposited On:30 Dec 2014 09:37
Last Modified:05 Apr 2016 18:39
Publisher:Springer
ISSN:1387-1307
Additional Information:The final publication is available at Springer via http://dx.doi.org/10.1007/s10877-014-9639-6
Publisher DOI:https://doi.org/10.1007/s10877-014-9639-6
PubMed ID:25388511
Permanent URL: https://doi.org/10.5167/uzh-102630

Download

[img]
Preview
Content: Accepted Version
Filetype: PDF
Size: 150kB
View at publisher

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.

Author Collaborations