Evaluation of accuracy of invasive and non-invasive blood pressure monitoring in relation to carotid artery pressure in anaesthetised ponies


Gent, Thomas C; Schwarz, Andrea; Hatz, Lea-Annina; Gozalo-Marcilla, Miguel; Stijn, Schauvliege; Gasthuys, Frank; Bettschart-Wolfensberger, Regula (2015). Evaluation of accuracy of invasive and non-invasive blood pressure monitoring in relation to carotid artery pressure in anaesthetised ponies. Pferdeheilkunde, (31):33-38.

Abstract

Invasive blood pressure measurement (IBP) using peripheral arteries is a commonly used technique in equine anaesthesia, although the accuracy has not been demonstrated. Non-invasive blood pressure monitoring (NIBP) may be indicated for field anesthesia, short procedures and foal anaesthesia. In the present report, the agreement of various IBP and NIBP measuring sites compared to carotid artery pressure was tested in anaesthetised experimental ponies. Six ponies were anaesthetised in lateral recumbency with sevoflurane and received either saline or dexmedetomidine boli followed by constant rate infusion (CRI). Invasive blood pressure measurements were obtained simultaneously from the carotid, facial and metatarsal arteries. NIBP measurements over both median arteries, metatarsal and middle coccygeal arteries were performed in random order. All blood pressure readings obtained were compared to carotid pressure by Bland-Altman analysis. Non-invasive blood pressure measurements had larger bias and poorer limits of agreement compared to IBP measurements. NIBP measurement from the coccygeal artery had the best repeatability and best limits of agreement of all NIBP positions and was not affected by the use of dexmedetomidine. The facial artery had smaller limits of agreement with dexmedetomidine treatment. There was no difference between the facial and metatarsal arteries during dexmedetomidine treatment in MAP and DAP. Systolic arterial pressures from the metatarsal artery showed larger bias and larger limits of agreement compared to facial artery. In conclusion, NIBP measurement from the tail artery is a reliable alternative to direct arterial blood pressure measurement. Both the metatarsal and facial arteries have acceptable agreement with carotid artery pressure so the choice can be based on the logistics of the procedure.

Invasive blood pressure measurement (IBP) using peripheral arteries is a commonly used technique in equine anaesthesia, although the accuracy has not been demonstrated. Non-invasive blood pressure monitoring (NIBP) may be indicated for field anesthesia, short procedures and foal anaesthesia. In the present report, the agreement of various IBP and NIBP measuring sites compared to carotid artery pressure was tested in anaesthetised experimental ponies. Six ponies were anaesthetised in lateral recumbency with sevoflurane and received either saline or dexmedetomidine boli followed by constant rate infusion (CRI). Invasive blood pressure measurements were obtained simultaneously from the carotid, facial and metatarsal arteries. NIBP measurements over both median arteries, metatarsal and middle coccygeal arteries were performed in random order. All blood pressure readings obtained were compared to carotid pressure by Bland-Altman analysis. Non-invasive blood pressure measurements had larger bias and poorer limits of agreement compared to IBP measurements. NIBP measurement from the coccygeal artery had the best repeatability and best limits of agreement of all NIBP positions and was not affected by the use of dexmedetomidine. The facial artery had smaller limits of agreement with dexmedetomidine treatment. There was no difference between the facial and metatarsal arteries during dexmedetomidine treatment in MAP and DAP. Systolic arterial pressures from the metatarsal artery showed larger bias and larger limits of agreement compared to facial artery. In conclusion, NIBP measurement from the tail artery is a reliable alternative to direct arterial blood pressure measurement. Both the metatarsal and facial arteries have acceptable agreement with carotid artery pressure so the choice can be based on the logistics of the procedure.

Citations

1 citation in Web of Science®
1 citation in Scopus®
Google Scholar™

Altmetrics

Downloads

10 downloads since deposited on 07 Jan 2015
6 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:05 Vetsuisse Faculty > Veterinary Clinic > Equine Department
?? 11061 ??
Dewey Decimal Classification:570 Life sciences; biology
630 Agriculture
Language:English
Date:2015
Deposited On:07 Jan 2015 09:55
Last Modified:20 Nov 2016 06:18
Publisher:Hippiatrika Verlag GmbH
ISSN:0177-7726
Free access at:Related URL. An embargo period may apply.
Publisher DOI:https://doi.org/10.21836/PEM20150105
Official URL:http://www.biblioserver.com/pferdeheilkunde-fundus/index.php?m=search&id=&ftype=data&q=20150105
Related URLs:http://www.hippiatrika.com/download.htm?id=en20150105
Permanent URL: https://doi.org/10.5167/uzh-103193

Download

[img]
Preview
Content: Published Version
Filetype: PDF
Size: 438kB
View at publisher

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.

Author Collaborations