Cardiac output measurement in children: comparison of Aesculon cardiac output monitor and thermodilution


Tomaske, M; Knirsch, W; Kretschmar, O; Woitzek, K; Balmer, C; Schmitz, A; Bauersfeld, U; Weiss, M (2008). Cardiac output measurement in children: comparison of Aesculon cardiac output monitor and thermodilution. British Journal of Anaesthesia, 100(4):517-520.

Abstract

BACKGROUND: We compared cardiac output (CO) measurements by the non-invasive electrical velocimetry (Aesculon) monitor with the pulmonary artery catheter (PAC) thermodilution method in children. METHODS: CO values using the Aesculon monitor and PAC thermodilution were simultaneously recorded during cardiac catheterization in children. Measurements were performed under general anaesthesia. To compare, three consecutive measurements for each patient within 3 min were obtained. The means of the three values were compared using simple regression and Bland-Altman analysis. Data were presented as mean (sd). A mean percentage of <30% was defined to indicate clinical useful reliability of the Aesculon monitor. RESULTS: A total of 50 patients with a median (range) age of 7.5 (0.5-16.5) yr were enrolled in the study. Mean CO values were 3.7 (1.5) litre min(-1) (PAC thermodilution) and 3.1 (1.7) litre min(-1) (Aesculon) monitor). Analysis for CO measurement showed a good correlation between the two methods (r=0.894; P<0.0001). The bias between the two methods was 0.66 litre min(-1) with a precision of 1.49 litre min(-1). The mean percentage error for CO measurements was 48.9% for the Aesculon monitor when compared with PAC thermodilution. CONCLUSIONS: Electrical velocimetry using the Aesculon monitor did not provide reliable CO values when compared with PAC thermodilution. Whether the Aesculon monitor can be used as a CO trend monitor has to be assessed by further investigations in patients with changing haemodynamics.

BACKGROUND: We compared cardiac output (CO) measurements by the non-invasive electrical velocimetry (Aesculon) monitor with the pulmonary artery catheter (PAC) thermodilution method in children. METHODS: CO values using the Aesculon monitor and PAC thermodilution were simultaneously recorded during cardiac catheterization in children. Measurements were performed under general anaesthesia. To compare, three consecutive measurements for each patient within 3 min were obtained. The means of the three values were compared using simple regression and Bland-Altman analysis. Data were presented as mean (sd). A mean percentage of <30% was defined to indicate clinical useful reliability of the Aesculon monitor. RESULTS: A total of 50 patients with a median (range) age of 7.5 (0.5-16.5) yr were enrolled in the study. Mean CO values were 3.7 (1.5) litre min(-1) (PAC thermodilution) and 3.1 (1.7) litre min(-1) (Aesculon) monitor). Analysis for CO measurement showed a good correlation between the two methods (r=0.894; P<0.0001). The bias between the two methods was 0.66 litre min(-1) with a precision of 1.49 litre min(-1). The mean percentage error for CO measurements was 48.9% for the Aesculon monitor when compared with PAC thermodilution. CONCLUSIONS: Electrical velocimetry using the Aesculon monitor did not provide reliable CO values when compared with PAC thermodilution. Whether the Aesculon monitor can be used as a CO trend monitor has to be assessed by further investigations in patients with changing haemodynamics.

Citations

26 citations in Web of Science®
29 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

86 downloads since deposited on 30 Jan 2009
14 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Children's Hospital Zurich > Clinic for Surgery
04 Faculty of Medicine > University Children's Hospital Zurich > Medical Clinic
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:2008
Deposited On:30 Jan 2009 21:55
Last Modified:05 Apr 2016 12:49
Publisher:Oxford University Press
ISSN:0007-0912
Publisher DOI:10.1093/bja/aen024
PubMed ID:18305081
Permanent URL: http://doi.org/10.5167/uzh-10325

Download

[img]
Preview
Filetype: PDF
Size: 1MB
View at publisher

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.

Author Collaborations