UZH-Logo

Maintenance Infos

Transcatheter aortic valve implantation of the direct flow medical aortic valve with minimal or no contrast


Latib, Azeem; Maisano, Francesco; Colombo, Antonio; Klugmann, Silvio; Low, Reginald; Smith, Thomas; Davidson, Charles; Harreld, John H; Bruschi, Giuseppe; DeMarco, Federico (2014). Transcatheter aortic valve implantation of the direct flow medical aortic valve with minimal or no contrast. Cardiovascular Revascularization Medicine, 15(4):252-257.

Abstract

The 18F Direct Flow Medical (DFM) THV has conformable sealing rings, which minimizes aortic regurgitation and permits full hemodynamic assessment of valve performance prior to permanent implantation. During the DISCOVER trial, three patients who were at risk for receiving contrast media, two due to severe CKD and one due to a recent hyperthyroid reaction to contrast, underwent DFM implantation under fluoroscopic and transesophageal guidance without aortography during either positioning or to confirm the final position. Valve positioning was based on the optimal angiographic projection as calculated by the pre-procedural multislice CT scan. Precise optimization of valve position was performed to minimize transvalve gradient and aortic regurgitation. Prior to final implantation, transvalve hemodynamics were assessed invasively and by TEE. The post-procedure mean gradients were 7, 10, 11mm Hg. The final AVA by echo was 1.70, 1.40 and 1.68cm(2). Total aortic regurgitation post-procedure was none or trace in all three patients. Total positioning and assessment of valve performance time was 4, 6, and 12minutes. Contrast was only used to confirm successful percutaneous closure of the femoral access site. The total contrast dose was 5, 8, 12cc. Baseline eGFR and creatinine was 28, 22, 74mL/min/1.73m(2) and 2.35, 2.98, and 1.03mg/dL, respectively. Renal function was unchanged post-procedure: eGFR=25, 35, and 96mL/min/1.73m(2) and creatinine=2.58, 1.99, and 1.03mg/dL, respectively. In conclusion, the DFM THV provides the ability to perform TAVI with minimal or no contrast. The precise and predictable implantation technique can be performed with fluoro and echo guidance.

The 18F Direct Flow Medical (DFM) THV has conformable sealing rings, which minimizes aortic regurgitation and permits full hemodynamic assessment of valve performance prior to permanent implantation. During the DISCOVER trial, three patients who were at risk for receiving contrast media, two due to severe CKD and one due to a recent hyperthyroid reaction to contrast, underwent DFM implantation under fluoroscopic and transesophageal guidance without aortography during either positioning or to confirm the final position. Valve positioning was based on the optimal angiographic projection as calculated by the pre-procedural multislice CT scan. Precise optimization of valve position was performed to minimize transvalve gradient and aortic regurgitation. Prior to final implantation, transvalve hemodynamics were assessed invasively and by TEE. The post-procedure mean gradients were 7, 10, 11mm Hg. The final AVA by echo was 1.70, 1.40 and 1.68cm(2). Total aortic regurgitation post-procedure was none or trace in all three patients. Total positioning and assessment of valve performance time was 4, 6, and 12minutes. Contrast was only used to confirm successful percutaneous closure of the femoral access site. The total contrast dose was 5, 8, 12cc. Baseline eGFR and creatinine was 28, 22, 74mL/min/1.73m(2) and 2.35, 2.98, and 1.03mg/dL, respectively. Renal function was unchanged post-procedure: eGFR=25, 35, and 96mL/min/1.73m(2) and creatinine=2.58, 1.99, and 1.03mg/dL, respectively. In conclusion, the DFM THV provides the ability to perform TAVI with minimal or no contrast. The precise and predictable implantation technique can be performed with fluoro and echo guidance.

Citations

Altmetrics

Downloads

0 downloads since deposited on 18 Feb 2015
0 downloads since 12 months

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Cardiovascular Surgery
Dewey Decimal Classification:610 Medicine & health
Language:German
Date:June 2014
Deposited On:18 Feb 2015 16:14
Last Modified:05 Apr 2016 18:56
Publisher:Elsevier
ISSN:1878-0938
Publisher DOI:https://doi.org/10.1016/j.carrev.2014.03.003
PubMed ID:24721585
Permanent URL: https://doi.org/10.5167/uzh-106643

Download

[img]
Content: Published Version
Filetype: PDF - Registered users only
Size: 936kB
View at publisher

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.