UZH-Logo

Maintenance Infos

Infection rate of emergency bolt-kit vs. non-emergency conventional implanted silver bearing external ventricular drainage catheters


Fichtner, Jens; Jilch, Astrid; Stieglitz, Lennart Henning; Beck, Jürgen; Raabe, Andreas; Z' Graggen, Werner J (2014). Infection rate of emergency bolt-kit vs. non-emergency conventional implanted silver bearing external ventricular drainage catheters. Clinical Neurology and Neurosurgery, 122:70-76.

Abstract

BACKGROUND Bolt-kit systems are increasingly used as an alternative to conventional external cerebrospinal fluid (CSF) drainage systems. Since 2009 we regularly utilize bolt-kit external ventricular drainage (EVD) systems with silver-bearing catheters inserted manually with a hand drill and skull screws for emergency ventriculostomy. For non-emergency situations, we use conventional ventriculostomy with subcutaneous tunneled silver-bearing catheters, performed in the operating room with a pneumatic drill. This retrospective analysis compared the two techniques in terms of infection rates.

METHODS 152 patients (aged 17-85 years, mean=55.4 years) were included in the final analysis; 95 received bolt-kit silver-bearing catheters and 57 received conventionally implanted silver-bearing catheters. The primary endpoint combined infection parameters: occurrence of positive CSF culture, colonization of catheter tips, or elevated CSF white blood cell counts (>4/μl). Secondary outcome parameters were presence of microorganisms in CSF or on catheter tips. Incidence of increased CSF cell counts and number of patients with catheter malposition were also compared.

RESULTS The primary outcome, defined as analysis of combined infection parameters (occurrence of either positive CSF culture, colonization of the catheter tips or raised CSF white blood cell counts >4/μl)was not significantly different between the groups (58.9% bolt-kit group vs. 63.2% conventionally implanted group, p=0.61, chi-square-test). The bolt-kit group was non-inferior and not superior to the conventional group (relative risk reduction of 6.7%; 90% confidence interval: -19.9% to 25.6%). Secondary outcomes showed no statistically significant difference in the incidence of microorganisms in CSF (2.1% bolt-kit vs. 5.3% conventionally implanted; p=0.30; chi-square-test).

CONCLUSIONS This analysis indicates that silver-bearing EVD catheters implanted with a bolt-kit system outside the operating room do not significantly elevate the risk of CSF infection as compared to conventional implant methods.

BACKGROUND Bolt-kit systems are increasingly used as an alternative to conventional external cerebrospinal fluid (CSF) drainage systems. Since 2009 we regularly utilize bolt-kit external ventricular drainage (EVD) systems with silver-bearing catheters inserted manually with a hand drill and skull screws for emergency ventriculostomy. For non-emergency situations, we use conventional ventriculostomy with subcutaneous tunneled silver-bearing catheters, performed in the operating room with a pneumatic drill. This retrospective analysis compared the two techniques in terms of infection rates.

METHODS 152 patients (aged 17-85 years, mean=55.4 years) were included in the final analysis; 95 received bolt-kit silver-bearing catheters and 57 received conventionally implanted silver-bearing catheters. The primary endpoint combined infection parameters: occurrence of positive CSF culture, colonization of catheter tips, or elevated CSF white blood cell counts (>4/μl). Secondary outcome parameters were presence of microorganisms in CSF or on catheter tips. Incidence of increased CSF cell counts and number of patients with catheter malposition were also compared.

RESULTS The primary outcome, defined as analysis of combined infection parameters (occurrence of either positive CSF culture, colonization of the catheter tips or raised CSF white blood cell counts >4/μl)was not significantly different between the groups (58.9% bolt-kit group vs. 63.2% conventionally implanted group, p=0.61, chi-square-test). The bolt-kit group was non-inferior and not superior to the conventional group (relative risk reduction of 6.7%; 90% confidence interval: -19.9% to 25.6%). Secondary outcomes showed no statistically significant difference in the incidence of microorganisms in CSF (2.1% bolt-kit vs. 5.3% conventionally implanted; p=0.30; chi-square-test).

CONCLUSIONS This analysis indicates that silver-bearing EVD catheters implanted with a bolt-kit system outside the operating room do not significantly elevate the risk of CSF infection as compared to conventional implant methods.

Altmetrics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Neurosurgery
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:1 July 2014
Deposited On:17 Feb 2015 17:04
Last Modified:05 Apr 2016 19:00
Publisher:Elsevier
ISSN:0303-8467
Publisher DOI:https://doi.org/10.1016/j.clineuro.2014.04.018
PubMed ID:24908221

Download

Full text not available from this repository.View at publisher

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.