Validation of a photoacoustic gas analyser for the measurement of functional residual capacity using multiple-breath inert gas washout


Gonem, Sherif; Singer, Florian; Corkill, Steven; Singapuri, Amisha; Siddiqui, Salman; Gustafsson, Per (2014). Validation of a photoacoustic gas analyser for the measurement of functional residual capacity using multiple-breath inert gas washout. Respiration, 87(6):462-468.

Abstract

BACKGROUND: The respiratory mass spectrometer is the current gold-standard technique for performing multiple-breath inert gas washout (MBW), but is expensive and lacks portability. A number of alternative techniques have recently been described.
OBJECTIVES: We aimed to validate, using an in vitro lung model, an open-circuit MBW system that utilises a portable photoacoustic gas analyser, with sulphur hexafluoride (SF6) as the inert tracer gas.
METHODS: An acrylic glass lung model was utilised to assess the accuracy of functional residual capacity (FRC) measurements derived from MBW. Measurements were performed in triplicate at 20 combinations of simulated FRC, tidal volume and respiratory rate. FRC measured using MBW (FRCmbw) was compared to FRC calculated from the known dimensions of the model (FRCcalc). MBW was also performed in 10 healthy subjects and 14 patients with asthma.
RESULTS: The MBW system measured FRC with high precision. The mean bias of FRCmbw with respect to FRCcalc was -0.4% (95% limits of agreement of -4.6 and 3.9%). The mean coefficient of variation of triplicate FRC measurements was 4.0% in vivo and 1.0% in vitro. MBW slightly underestimated low lung volumes and overestimated high lung volumes, but this did not cause a significant error in lung clearance index except at lung volumes below 1,500 ml.
CONCLUSIONS: The open-circuit MBW system utilising SF6 as the inert tracer gas and a photoacoustic gas analyser is both accurate and repeatable within the adult range of lung volumes. Further modifications would be required before its use in young children or infants.

BACKGROUND: The respiratory mass spectrometer is the current gold-standard technique for performing multiple-breath inert gas washout (MBW), but is expensive and lacks portability. A number of alternative techniques have recently been described.
OBJECTIVES: We aimed to validate, using an in vitro lung model, an open-circuit MBW system that utilises a portable photoacoustic gas analyser, with sulphur hexafluoride (SF6) as the inert tracer gas.
METHODS: An acrylic glass lung model was utilised to assess the accuracy of functional residual capacity (FRC) measurements derived from MBW. Measurements were performed in triplicate at 20 combinations of simulated FRC, tidal volume and respiratory rate. FRC measured using MBW (FRCmbw) was compared to FRC calculated from the known dimensions of the model (FRCcalc). MBW was also performed in 10 healthy subjects and 14 patients with asthma.
RESULTS: The MBW system measured FRC with high precision. The mean bias of FRCmbw with respect to FRCcalc was -0.4% (95% limits of agreement of -4.6 and 3.9%). The mean coefficient of variation of triplicate FRC measurements was 4.0% in vivo and 1.0% in vitro. MBW slightly underestimated low lung volumes and overestimated high lung volumes, but this did not cause a significant error in lung clearance index except at lung volumes below 1,500 ml.
CONCLUSIONS: The open-circuit MBW system utilising SF6 as the inert tracer gas and a photoacoustic gas analyser is both accurate and repeatable within the adult range of lung volumes. Further modifications would be required before its use in young children or infants.

Citations

6 citations in Web of Science®
7 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

2 downloads since deposited on 12 Feb 2015
2 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Children's Hospital Zurich > Medical Clinic
Dewey Decimal Classification:610 Medicine & health
Date:2014
Deposited On:12 Feb 2015 14:34
Last Modified:03 Jun 2016 10:22
Publisher:Karger
ISSN:0025-7931
Publisher DOI:https://doi.org/10.1159/000357786
PubMed ID:24525760
Permanent URL: https://doi.org/10.5167/uzh-108082

Download

[img]
Preview
Content: Published Version
Filetype: PDF
Size: 311kB
View at publisher

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.