Head to head evaluation of the analytical performance of two commercial methotrexate immunoassays and comparison with liquid chromatography-mass spectrometry and the former fluorescence polarization immunoassay


Günther, Viola; Mueller, Daniel; von Eckardstein, Arnold; Saleh, Lanja (2015). Head to head evaluation of the analytical performance of two commercial methotrexate immunoassays and comparison with liquid chromatography-mass spectrometry and the former fluorescence polarization immunoassay. Clinical Chemistry and Laboratory Medicine:online.

Abstract

BACKGROUND: onitoring of plasma drug levels is mandatory in patients receiving high-dose methotrexate. This study evaluated the analytical performance of the novel Architect and the established ARK™ methotrexate immunoassay (running on the Roche Cobas© c502 analyzer) in comparison with liquid chromatography-mass spectrometry (LC-MS) and the TDx/TDxFLx Methotrexate II assay.
METHODS: Imprecision and linearity were verified for the Architect and ARK assay according to CLSI EP15-A3 and EP6-A guidelines, respectively. The reported limit of quantitation (0.04 μmol/L) was tested for both assays according to the CLSI EP17-A2 guideline. Correlation and agreement between the different assays were evaluated using residual plasma samples (n=153).
RESULTS: Total imprecision was <6.3% and <9.5% for the Architect and ARK assay, respectively. The claimed linearity and limit of quantitation were confirmed for the Architect assay. For the ARK assay, imprecision at the limit of quantitation was <18% with a positive bias resulting in a high total error up to 58%, and hence the linearity could not be confirmed. Both assays showed strong correlations with the TDX assay and LC-MS but a positive bias of 12.2% and 20.5% in comparison to LC-MS for the Architect and ARK assay, respectively. For the ARK assay this bias increased dramatically for samples with concentrations towards the limit of quantitation.
CONCLUSIONS: The Architect assay is suitable for monitoring plasma methotrexate, but the ARK assay showed unsatisfactory performance in the analysis of low concentrated samples. Unlike the TDX assay, both assays require manual dilution of samples at higher concentrations, which delays sample processing in clinical routine.

Abstract

BACKGROUND: onitoring of plasma drug levels is mandatory in patients receiving high-dose methotrexate. This study evaluated the analytical performance of the novel Architect and the established ARK™ methotrexate immunoassay (running on the Roche Cobas© c502 analyzer) in comparison with liquid chromatography-mass spectrometry (LC-MS) and the TDx/TDxFLx Methotrexate II assay.
METHODS: Imprecision and linearity were verified for the Architect and ARK assay according to CLSI EP15-A3 and EP6-A guidelines, respectively. The reported limit of quantitation (0.04 μmol/L) was tested for both assays according to the CLSI EP17-A2 guideline. Correlation and agreement between the different assays were evaluated using residual plasma samples (n=153).
RESULTS: Total imprecision was <6.3% and <9.5% for the Architect and ARK assay, respectively. The claimed linearity and limit of quantitation were confirmed for the Architect assay. For the ARK assay, imprecision at the limit of quantitation was <18% with a positive bias resulting in a high total error up to 58%, and hence the linearity could not be confirmed. Both assays showed strong correlations with the TDX assay and LC-MS but a positive bias of 12.2% and 20.5% in comparison to LC-MS for the Architect and ARK assay, respectively. For the ARK assay this bias increased dramatically for samples with concentrations towards the limit of quantitation.
CONCLUSIONS: The Architect assay is suitable for monitoring plasma methotrexate, but the ARK assay showed unsatisfactory performance in the analysis of low concentrated samples. Unlike the TDX assay, both assays require manual dilution of samples at higher concentrations, which delays sample processing in clinical routine.

Altmetrics

Downloads

0 downloads since deposited on 08 Feb 2016
0 downloads since 12 months

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Institute of Clinical Chemistry
Dewey Decimal Classification:610 Medicine & health
540 Chemistry
Language:English
Date:10 October 2015
Deposited On:08 Feb 2016 14:09
Last Modified:11 Oct 2016 00:00
Publisher:De Gruyter
ISSN:1434-6621
Publisher DOI:https://doi.org/10.1515/cclm-2015-0578
PubMed ID:26457783

Download

[img]
Preview
Content: Published Version
Filetype: PDF
Size: 506kB
View at publisher

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.

Author Collaborations