One-year patency control and risk analysis of eSVS®-mesh-supported coronary saphenous vein grafts


Inderbitzin, Devdas T; Bremerich, Jens; Matt, Peter; Grapow, Martin T R; Eckstein, Friedrich S; Reuthebuch, Oliver (2015). One-year patency control and risk analysis of eSVS®-mesh-supported coronary saphenous vein grafts. Journal of Cardiothoracic Surgery, 10(108):online.

Abstract

BACKGROUND: The eSVS® external venous nitinol mesh (Kips Bay Medical, Minneapolis, USA) was designed to improve long-term patency of coronary saphenous vein grafts (SVG) by preventing pressure-induced wall stress and reactive neo-intimal hyperplasia. We present one-year-patency rates of meshed SVGs assessed by coronary computed tomographic angiography (cCTA).
PATIENTS AND METHODS: Data from consecutive patients receiving an eSVS® meshed coronary bypass SVG from 06/2010 to 06/2011 were prospectively collected and analysed post-hoc. Patient characteristics, coronary artery disease, SVG quality, surgery (including number of anastomoses and transit time flow-measurement: TTFM), postoperative course and graft patency by cCTA were recorded. Potential risk factors for meshed graft occlusion were evaluated.
RESULTS: 22 patients received an eSVS® mesh (18 isolated CABG, 4 combined with aortic valve replacement). Three patients died prior to the one-year follow-up and were excluded. All 19 surviving patients (mean age 70.4 ± 9.5 years, 3 female) completed a cCTA of all grafts at 12 ± 0.1 months after surgery including 21 meshed SVGs (33 distal anastomoses), 7 unmeshed SVGs (13 distal anastomoses) and 22 arterial grafts (30 distal anastomoses). Mesh application was safe with patent grafts (by intraoperative TTFM) and perioperative course uneventful in all patients. The average graft/anastomosis number per patient was 2.6 ± 0.5/3.7 ± 0.8. Patency was unrestricted in all arterial and unmeshed SVGs (cCTA). Meshed SVG patency was 85 % (n = 28/33) for distal anastomoses and 76 % (n = 16/21) among meshed SVGs. Four SVGs with single distal anastomosis to the right coronary were completely occluded. One sequential graft to the left coronary was occluded between proximal and first distal anastomosis (see Fig. 1). Patency was independent of target site, coronary run-off, SVG quality and sequential distal grafting. All patients were asymptomatic.
CONCLUSIONS: The overall one-year patency rate of eSVS® meshed SVGs/anastomoses was 76 %/85 %. Surgical implantation is safe independently of target site, run-off, vein quality and sequential distal anastomoses. However, graft patency of meshed veins (76 %) was inferior to non-meshed (100 %) or arterial grafts (100 %). Thus our mid-term data do not sustain the concept of improving vein graft patency by external reinforcing with the eSVS® mesh. Further long-term follow-up is warranted.

Abstract

BACKGROUND: The eSVS® external venous nitinol mesh (Kips Bay Medical, Minneapolis, USA) was designed to improve long-term patency of coronary saphenous vein grafts (SVG) by preventing pressure-induced wall stress and reactive neo-intimal hyperplasia. We present one-year-patency rates of meshed SVGs assessed by coronary computed tomographic angiography (cCTA).
PATIENTS AND METHODS: Data from consecutive patients receiving an eSVS® meshed coronary bypass SVG from 06/2010 to 06/2011 were prospectively collected and analysed post-hoc. Patient characteristics, coronary artery disease, SVG quality, surgery (including number of anastomoses and transit time flow-measurement: TTFM), postoperative course and graft patency by cCTA were recorded. Potential risk factors for meshed graft occlusion were evaluated.
RESULTS: 22 patients received an eSVS® mesh (18 isolated CABG, 4 combined with aortic valve replacement). Three patients died prior to the one-year follow-up and were excluded. All 19 surviving patients (mean age 70.4 ± 9.5 years, 3 female) completed a cCTA of all grafts at 12 ± 0.1 months after surgery including 21 meshed SVGs (33 distal anastomoses), 7 unmeshed SVGs (13 distal anastomoses) and 22 arterial grafts (30 distal anastomoses). Mesh application was safe with patent grafts (by intraoperative TTFM) and perioperative course uneventful in all patients. The average graft/anastomosis number per patient was 2.6 ± 0.5/3.7 ± 0.8. Patency was unrestricted in all arterial and unmeshed SVGs (cCTA). Meshed SVG patency was 85 % (n = 28/33) for distal anastomoses and 76 % (n = 16/21) among meshed SVGs. Four SVGs with single distal anastomosis to the right coronary were completely occluded. One sequential graft to the left coronary was occluded between proximal and first distal anastomosis (see Fig. 1). Patency was independent of target site, coronary run-off, SVG quality and sequential distal grafting. All patients were asymptomatic.
CONCLUSIONS: The overall one-year patency rate of eSVS® meshed SVGs/anastomoses was 76 %/85 %. Surgical implantation is safe independently of target site, run-off, vein quality and sequential distal anastomoses. However, graft patency of meshed veins (76 %) was inferior to non-meshed (100 %) or arterial grafts (100 %). Thus our mid-term data do not sustain the concept of improving vein graft patency by external reinforcing with the eSVS® mesh. Further long-term follow-up is warranted.

Altmetrics

Downloads

3 downloads since deposited on 21 Jan 2016
3 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Cardiovascular Surgery
Dewey Decimal Classification:610 Medicine & health
Language:German
Date:2015
Deposited On:21 Jan 2016 13:17
Last Modified:05 Apr 2016 19:57
Publisher:BioMed Central
ISSN:1749-8090
Free access at:Publisher DOI. An embargo period may apply.
Publisher DOI:https://doi.org/10.1186/s13019-015-0293-y
PubMed ID:26253565

Download

[img]
Preview
Content: Published Version
Filetype: PDF
Size: 1MB
View at publisher
Licence: Creative Commons: Attribution 4.0 International (CC BY 4.0)

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.

Author Collaborations