Comparison of near-infrared oximeters in a liquid optical phantom with varying intralipid and blood content


Kleiser, S; Hyttel-Sorensen, S; Greisen, G; Wolf, M (2016). Comparison of near-infrared oximeters in a liquid optical phantom with varying intralipid and blood content. Advances in Experimental Medicine and Biology, 876:413-418.

Abstract

The interpretation of cerebral tissue oxygen saturation values (StO2) in clinical settings is currently complicated by the use of different near-infrared spectrophotometry (NIRS) devices producing different StO2 values for the same oxygenation due to differences in the algorithms and technical aspects. The aim was to investigate the effect of changes in scattering and absorption on the StO2 of different NIRS devices in a liquid optical phantom. We compared three continuous-wave (CW) with a frequency domain (FD) NIRS device. Responsiveness to oxygenation changes was only slightly altered by different intralipid (IL) concentrations. However, alterations in haematocrit (htc) showed a strong effect: increased htc led to a 20-35% increased response of all CW devices compared to the FD device, probably due to differences in algorithms regarding the water concentration.

Abstract

The interpretation of cerebral tissue oxygen saturation values (StO2) in clinical settings is currently complicated by the use of different near-infrared spectrophotometry (NIRS) devices producing different StO2 values for the same oxygenation due to differences in the algorithms and technical aspects. The aim was to investigate the effect of changes in scattering and absorption on the StO2 of different NIRS devices in a liquid optical phantom. We compared three continuous-wave (CW) with a frequency domain (FD) NIRS device. Responsiveness to oxygenation changes was only slightly altered by different intralipid (IL) concentrations. However, alterations in haematocrit (htc) showed a strong effect: increased htc led to a 20-35% increased response of all CW devices compared to the FD device, probably due to differences in algorithms regarding the water concentration.

Citations

1 citation in Web of Science®
1 citation in Scopus®
Google Scholar™

Altmetrics

Downloads

1 download since deposited on 27 Sep 2016
1 download since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Neonatology
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:2016
Deposited On:27 Sep 2016 08:16
Last Modified:27 Sep 2016 08:16
Publisher:Springer
ISSN:0065-2598
Publisher DOI:https://doi.org/10.1007/978-1-4939-3023-4_52
PubMed ID:26782240

Download

[img]
Content: Published Version
Filetype: PDF - Registered users only
Size: 262kB
View at publisher

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.

Author Collaborations