Influence of inflow cannula length in axial-flow pumps on neurologic adverse event rate: results from a multi-center analysis


Schmid, C; Jurmann, M; Birnbaum, D; Colombo, T; Falk, V; Feltrin, G; Garatti, A; Genoni, M; Gerosa, G; Göttel, P; Gummert, J; Halfmann, R; Hammel, D; Hennig, E; Kaufmann, F; Lanfranconi, M; Meyns, B; Mohr, F; Müller, J; Nikolov, D; Rucinskas, K; Scheld, H H; Schmid, F X; Schneider, M; Sirvydis, V; Tandler, R; Vitali, E; Vlasselaers, D; Weyand, M; Wilhelm, M J; Hetzer, R (2008). Influence of inflow cannula length in axial-flow pumps on neurologic adverse event rate: results from a multi-center analysis. Journal of Heart and Lung Transplantation, 27(3):253-260.

Abstract

BACKGROUND: The application of axial-flow pumps in patients with end-stage heart failure reveals a significantly reduced infectious complication rate as compared with rates observed with pulsatile devices. The remaining adverse event rate relates mainly to thromboembolic complications with neurologic consequences. We investigated the dependence of the neurologic adverse event rate on the length of the inflow cannula. METHODS: A total of 216 consecutive patients with an axial-flow pump (INCOR; Berlin Heart GmbH, Berlin, Germany) were included in a retrospective multi-center analysis. In 138 patients, a short inflow cannula (24-mm tip length into the left ventricle), and in 78 patients a long inflow cannula (tip length 34 mm) was applied. RESULTS: Patients with a long inflow cannula (LC) demonstrated a better survival rate than those with a short inflow cannula (SC) at the end of the observation period (LC, 63.4%; SC, 52.9%; p = 0.05). The thromboembolic adverse event rate was also significantly lower. Only 3 of the 78 patients (3.8%) with an LC had a thromboembolic adverse event (thromboembolic events per patient-year = 0.11) as compared with 32 (23.2%) of SC patients (thromboembolic events per patient-year = 0.50, p < 0.001). CONCLUSIONS: Patients with a long inflow cannula had a better survival rate and a lower incidence of cerebrovascular adverse events than patients with a short inflow cannula.

BACKGROUND: The application of axial-flow pumps in patients with end-stage heart failure reveals a significantly reduced infectious complication rate as compared with rates observed with pulsatile devices. The remaining adverse event rate relates mainly to thromboembolic complications with neurologic consequences. We investigated the dependence of the neurologic adverse event rate on the length of the inflow cannula. METHODS: A total of 216 consecutive patients with an axial-flow pump (INCOR; Berlin Heart GmbH, Berlin, Germany) were included in a retrospective multi-center analysis. In 138 patients, a short inflow cannula (24-mm tip length into the left ventricle), and in 78 patients a long inflow cannula (tip length 34 mm) was applied. RESULTS: Patients with a long inflow cannula (LC) demonstrated a better survival rate than those with a short inflow cannula (SC) at the end of the observation period (LC, 63.4%; SC, 52.9%; p = 0.05). The thromboembolic adverse event rate was also significantly lower. Only 3 of the 78 patients (3.8%) with an LC had a thromboembolic adverse event (thromboembolic events per patient-year = 0.11) as compared with 32 (23.2%) of SC patients (thromboembolic events per patient-year = 0.50, p < 0.001). CONCLUSIONS: Patients with a long inflow cannula had a better survival rate and a lower incidence of cerebrovascular adverse events than patients with a short inflow cannula.

Citations

19 citations in Web of Science®
25 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

1 download since deposited on 12 Feb 2009
0 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Cardiovascular Surgery
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:2008
Deposited On:12 Feb 2009 07:47
Last Modified:05 Apr 2016 12:59
Publisher:Elsevier
ISSN:1053-2498
Publisher DOI:10.1016/j.healun.2007.12.007
PubMed ID:18342745
Permanent URL: http://doi.org/10.5167/uzh-13448

Download

[img]Filetype: PDF - Registered users only
Size: 1MB
View at publisher

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.

Author Collaborations