Non-invasive central venous pressure measurement by compression ultrasound-A step into real life


Thalhammer, C; Siegemund, M; Aschwanden, M; Gassmann, M; Baumann, U A; Jaeger, K A; Imfeld, S (2009). Non-invasive central venous pressure measurement by compression ultrasound-A step into real life. Resuscitation, 80(10):1130-1136.

Abstract

Information on central venous pressure (CVP) is helpful in clinical situations like cardiac failure and sepsis. Compression ultrasound (CU) of a forearm vein has been shown to be a reliable method for CVP measurement when assessed by experienced personal under study conditions. To test the hypothesis, that CU can be reliably used for CVP measurement after a training program and using a portable ultrasound system, we investigated feasibility and accuracy of CU, comparing the results of vascular experts vs. trainees as well as high-end vs. a portable ultrasound system. METHODS: CU with non-invasive CVP measurement (CVP(ni)) was performed by four investigators in 50 patients of an intensive care unit and compared with invasive CVP measurement (CVP(i)). RESULTS: Feasibility was between 88 and 92% in the different investigator groups. CVP(ni) measurement showed a significant linear correlation (r=0.58-0.68; p<0.001) with CVP(i) in all groups. Mean difference between CVP(i) and CVP(ni) was 2.4+/-3.1mmHg and similar between the investigators. No differences were observed between measurements done by vascular experts and trainees, as well as between high-end and portable ultrasound systems. Further we demonstrated, that CU is able to detect changes in CVP during the respiratory cycle. CONCLUSIONS: After a training program CU can be used by non-vascular clinician for reliable CVP measurement with good quality portable systems. Furthermore, respiratory changes in CVP are detectable by CU. These data suggest CU to be an efficient method for measuring CVP under real life conditions.

Information on central venous pressure (CVP) is helpful in clinical situations like cardiac failure and sepsis. Compression ultrasound (CU) of a forearm vein has been shown to be a reliable method for CVP measurement when assessed by experienced personal under study conditions. To test the hypothesis, that CU can be reliably used for CVP measurement after a training program and using a portable ultrasound system, we investigated feasibility and accuracy of CU, comparing the results of vascular experts vs. trainees as well as high-end vs. a portable ultrasound system. METHODS: CU with non-invasive CVP measurement (CVP(ni)) was performed by four investigators in 50 patients of an intensive care unit and compared with invasive CVP measurement (CVP(i)). RESULTS: Feasibility was between 88 and 92% in the different investigator groups. CVP(ni) measurement showed a significant linear correlation (r=0.58-0.68; p<0.001) with CVP(i) in all groups. Mean difference between CVP(i) and CVP(ni) was 2.4+/-3.1mmHg and similar between the investigators. No differences were observed between measurements done by vascular experts and trainees, as well as between high-end and portable ultrasound systems. Further we demonstrated, that CU is able to detect changes in CVP during the respiratory cycle. CONCLUSIONS: After a training program CU can be used by non-vascular clinician for reliable CVP measurement with good quality portable systems. Furthermore, respiratory changes in CVP are detectable by CU. These data suggest CU to be an efficient method for measuring CVP under real life conditions.

Citations

9 citations in Web of Science®
11 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

290 downloads since deposited on 23 Sep 2009
30 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Angiology
Dewey Decimal Classification:610 Medicine & health
Date:October 2009
Deposited On:23 Sep 2009 15:54
Last Modified:05 Apr 2016 13:20
Publisher:Elsevier
ISSN:0300-9572
Publisher DOI:10.1016/j.resuscitation.2009.06.027
PubMed ID:19632026
Permanent URL: http://doi.org/10.5167/uzh-20719

Download

[img]
Preview
Content: Accepted Version
Filetype: PDF
Size: 1MB
View at publisher

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.

Author Collaborations