Quick Search:

uzh logo
Browse by:
bullet
bullet
bullet
bullet

Zurich Open Repository and Archive

Permanent URL to this publication: http://dx.doi.org/10.5167/uzh-24038

Schirra, C O; Weiss, S; Krueger, S; Pedersen, S F; Razavi, R; Schaeffter, T; Kozerke, S (2009). Toward true 3D visualization of active catheters using compressed sensing. Magnetic Resonance in Medicine, 62(2):341-347.

[img] PDF - Registered users only
1MB

View at publisher

Abstract

A crucial requirement in MR-guided interventions is the visualization of catheter devices in real time. However, true 3D visualization of the full length of catheters has hitherto been impossible given scan time constraints. Compressed sensing (CS) has recently been proposed as a method to accelerate MR imaging of sparse objects. Images acquired with active interventional devices exhibit a high CNR and are inherently sparse, therefore rendering CS ideally suited for accelerating data acquisition.
A framework for true visualization of active catheters
in 3D is proposed employing CS to gain high undersampling
factors making real-time applications feasible. Constraints are introduced taking into account prior knowledge of catheter geometry and catheter motion over time to improve and accelerate image reconstruction. The potential of the method is demonstrated using computer simulations and phantom experiments and in vivo feasibility is demonstrated in a pig experiment.

Citations

Altmetrics

Downloads

1 download since deposited on 02 Dec 2009
0 downloads since 12 months

Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > Institute of Biomedical Engineering
DDC:170 Ethics
610 Medicine & health
Language:English
Date:August 2009
Deposited On:02 Dec 2009 14:42
Last Modified:23 Nov 2012 15:57
Publisher:Wiley-Blackwell
ISSN:0740-3194
Publisher DOI:10.1002/mrm.22001

Users (please log in): suggest update or correction for this item

Repository Staff Only: item control page