Permanent URL to this publication: http://dx.doi.org/10.5167/uzh-24044
Frauenrath, T; Hezel, F; Heinrichs, U; Kozerke, S; Utting, J F; Kob, M; Butenweg, C; Boesiger, P; Niendorf, T (2009). Feasibility of cardiac gating free of interference with electro-magnetic fields at 1.5 Tesla, 3.0 Tesla and 7.0 Tesla using an MR-stethoscope. Investigative Radiology, 44(9):539-547.
| PDF - Registered users only 3313Kb |
Abstract
OBJECTIVES: To circumvent the challenges of conventional electrocardiographic (ECG)-gating by examining the efficacy of an MR stethoscope, which offers (i) no risk of high voltage induction or patient burns, (ii) immunity to electromagnetic interference, (iii) suitability for all magnetic field strengths, and (iv) patient comfort together with ease of use for the pursuit of reliable and safe (ultra)high field cardiac gated magnetic resonance imaging (MRI). MATERIALS AND METHODS: The acoustic gating device consists of 3 main components: an acoustic sensor, a signal processing unit, and a coupler unit to the MRI system. Signal conditioning and conversion are conducted outside the 0.5 mT line using dedicated electronic circuits. The final waveform is delivered to the internal physiological signal controller circuitry of a clinical MR scanner. Cardiovascular MRI was performed of normal volunteers (n = 17) on 1.5 T, 3.0 T and 7.0 T whole body MR systems. Black blood imaging, 2D CINE imaging, 3D phase contrast MR angiography, and myocardial T2* mapping were carried out. RESULTS: The MR-stethoscope provided cardiograms at 1.5 T, 3.0 T and 7.0 T free of interference from electromagnetic fields and magneto-hydrodynamic effects. In comparison, ECG waveforms were susceptible to T-wave elevation and other distortions, which were more pronounced at higher fields. Acoustically gated black blood imaging at 1.5 T and 3.0 T provided image quality comparable with or even superior to that obtained from the ECG-gated approach. In the case of correct R-wave recognition, ECG-gated 2D CINE SSFP imaging was found to be immune to cardiac motion effects -even at 3.0 T. However, ECG-gated 2D SSFP CINE imaging was prone to cardiac motion artifacts if R-wave mis-registration occurred because of T-wave elevation. Acoustically gated 3D PCMRA at 1.5 T, 3.0 T and 7.0 T resulted in images free of blood pulsation artifacts because the acoustic gating approach provided cardiac signal traces free of interference with electromagnetic fields or magneto-hydrodynamic effects even at 7.0 Tesla. Severe ECG-trace distortions and T-wave elevations occurred at 3.0 T and 7.0 T. Acoustically cardiac gated T2* mapping at 3.0 T yielded a T2* value of 22.3 +/- 4.8 ms for the inferoseptal myocardium. CONCLUSIONS: The proposed MR-stethoscope presents a promising alternative to currently available techniques for cardiac gating of (ultra)high field MRI. Its intrinsic insensitivity to interference from electromagnetic fields renders it suitable for clinical imaging because of its excellent trigger reliability, even at 7.0 Tesla.
| Item Type: | Journal Article, refereed, original work |
|---|---|
| Communities & Collections: | 04 Faculty of Medicine > Institute of Biomedical Engineering |
| DDC: | 170 Ethics 610 Medicine & health |
| Language: | English |
| Date: | 2009 |
| Deposited On: | 18 Nov 2009 11:39 |
| Last Modified: | 09 Jul 2012 06:00 |
| Publisher: | Lippincott Wiliams & Wilkins |
| ISSN: | 0020-9996 |
| Publisher DOI: | 10.1097/RLI.0b013e3181b4c15e |
| PubMed ID: | 19652614 |
| WoS Citation Count: | 26 |
Users (please log in): suggest update or correction for this item
Repository Staff Only: item control page