Kontinuierliche Augeninnendruckmessung : Erste Ergebnisse einer drucksensitiven Kontaktlinse


Hediger, A; Kniestedt, C; Zweifel, S; Knecht, P; Funk, J; Kanngiesser, H (2009). Kontinuierliche Augeninnendruckmessung : Erste Ergebnisse einer drucksensitiven Kontaktlinse. Der Ophthalmologe, 106(12):1111-1115.

Abstract

A prototype of a pressure-sensitive contact lens (CL-DCT) based on dynamic contour tonometry (DCT) has been developed for continuous intraocular pressure (IOP) monitoring. The purpose of the present study was to assess the clinical applicability of this CL-DCT. MATERIAL AND METHODS: In 24 healthy subjects continuous IOP recording with the CL-DCT for 100 s was accomplished including IOP monitoring during 2 consecutive Valsalva manoeuvres. The measurement procedure was repeated after 5 min on the same eye and compared with the initial measurement. RESULTS: Continuous pressure curves were recorded in 20 subjects. The CL-DCT revealed an individual increase in IOP from 0.74 to 8.26 mmHg during the Valsalva manoeuvres. Comparison of the 2 consecutive IOP measurements yielded a mean deviation of +/-0.4 mmHg. CONCLUSIONS: CL-DCT allows non-invasive and continuous measurements of IOP. The measured values are comparable to the expected ones. Further studies are necessary to compare the measurement accuracy of CL-DCT with that of slit lamp adapted DCT (SL-DCT).

A prototype of a pressure-sensitive contact lens (CL-DCT) based on dynamic contour tonometry (DCT) has been developed for continuous intraocular pressure (IOP) monitoring. The purpose of the present study was to assess the clinical applicability of this CL-DCT. MATERIAL AND METHODS: In 24 healthy subjects continuous IOP recording with the CL-DCT for 100 s was accomplished including IOP monitoring during 2 consecutive Valsalva manoeuvres. The measurement procedure was repeated after 5 min on the same eye and compared with the initial measurement. RESULTS: Continuous pressure curves were recorded in 20 subjects. The CL-DCT revealed an individual increase in IOP from 0.74 to 8.26 mmHg during the Valsalva manoeuvres. Comparison of the 2 consecutive IOP measurements yielded a mean deviation of +/-0.4 mmHg. CONCLUSIONS: CL-DCT allows non-invasive and continuous measurements of IOP. The measured values are comparable to the expected ones. Further studies are necessary to compare the measurement accuracy of CL-DCT with that of slit lamp adapted DCT (SL-DCT).

Citations

7 citations in Web of Science®
10 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

331 downloads since deposited on 25 Nov 2009
35 downloads since 12 months
Detailed statistics

Additional indexing

Other titles:Continuous intraocular pressure measurement : First results with a pressure-sensitive contact lens
Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Ophthalmology Clinic
Dewey Decimal Classification:610 Medicine & health
Language:German
Date:12 September 2009
Deposited On:25 Nov 2009 08:47
Last Modified:05 Apr 2016 13:34
Publisher:Springer
ISSN:0941-293X
Additional Information:The original publication is available at www.springerlink.com
Publisher DOI:https://doi.org/10.1007/s00347-009-1919-z
PubMed ID:19756646
Permanent URL: https://doi.org/10.5167/uzh-24427

Download

[img]
Filetype: PDF - Registered users only
Size: 585kB
View at publisher
[img]
Preview
Content: Accepted Version
Filetype: PDF
Size: 2MB

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.

Author Collaborations