Evaluation of high-definition and conventional oscillometric blood pressure measurement in anaesthetised dogs using ACVIM guidelines


Wernick, M B; Doherr, M; Howard, J; Francey, T (2010). Evaluation of high-definition and conventional oscillometric blood pressure measurement in anaesthetised dogs using ACVIM guidelines. Journal of Small Animal Practice, 51(6):318-324.

Abstract

OBJECTIVES: To evaluate high-definition and conventional oscillometryin comparison with direct blood pressure measurements in anaesthetised dogs.
METHODS: Eight simultaneous readings for systolic, diastolic and mean pressure were obtained directly and with each of two devices in nine anaesthetised dogs. Measurement procedure and validation were based on the 2007 ACVIM guidelines.
RESULTS: Sixty-three simultaneous readings were evaluated for each device and direct measurements. The mean differences (bias) to direct values were within 10 mmHg for both devices although bias for systolic and diastolic blood pressures was higher for Memodiagnostic. The standard deviations of differences (precision) were within 15 mmHg for Dinamap but exceeded for Memodiagnostic. Correlation coefficients were higher for Dinamap than Memodiagnostic but both failed to reach a correlation of 0·9. Over 50% of values lay within 10 mmHg of direct measures for both devices, but this percentage was greater for Dinamap than Memodiagnostic. Over 80% of values lay within 20 mmHg of direct measures for Dinamap but not for Memodiagnostic.
CLINICAL SIGNIFICANCE: Both devices failed to meet ACVIM guideline validation. However, Dinamap only failed with regards to correlation. Memodiagnostic failed on several requirements, and based on poor correlation, accuracy and precision, this device cannot be currently recommended for dogs under anaesthesia.

OBJECTIVES: To evaluate high-definition and conventional oscillometryin comparison with direct blood pressure measurements in anaesthetised dogs.
METHODS: Eight simultaneous readings for systolic, diastolic and mean pressure were obtained directly and with each of two devices in nine anaesthetised dogs. Measurement procedure and validation were based on the 2007 ACVIM guidelines.
RESULTS: Sixty-three simultaneous readings were evaluated for each device and direct measurements. The mean differences (bias) to direct values were within 10 mmHg for both devices although bias for systolic and diastolic blood pressures was higher for Memodiagnostic. The standard deviations of differences (precision) were within 15 mmHg for Dinamap but exceeded for Memodiagnostic. Correlation coefficients were higher for Dinamap than Memodiagnostic but both failed to reach a correlation of 0·9. Over 50% of values lay within 10 mmHg of direct measures for both devices, but this percentage was greater for Dinamap than Memodiagnostic. Over 80% of values lay within 20 mmHg of direct measures for Dinamap but not for Memodiagnostic.
CLINICAL SIGNIFICANCE: Both devices failed to meet ACVIM guideline validation. However, Dinamap only failed with regards to correlation. Memodiagnostic failed on several requirements, and based on poor correlation, accuracy and precision, this device cannot be currently recommended for dogs under anaesthesia.

Citations

10 citations in Web of Science®
13 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

56 downloads since deposited on 05 Jul 2010
0 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:05 Vetsuisse Faculty > Veterinary Clinic > Department of Small Animals
Dewey Decimal Classification:570 Life sciences; biology
630 Agriculture
Language:English
Date:June 2010
Deposited On:05 Jul 2010 14:53
Last Modified:05 Apr 2016 14:10
Publisher:Wiley-Blackwell
ISSN:0022-4510
Publisher DOI:10.1111/j.1748-5827.2010.00938.x
Permanent URL: http://doi.org/10.5167/uzh-34547

Download

[img]
Filetype: PDF - Registered users only
Size: 542kB
View at publisher
[img]
Preview
Content: Accepted Version
Filetype: PDF
Size: 1MB

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.

Author Collaborations