Low kilovoltage cardiac dual-source CT: attenuation, noise, and radiation dose


Leschka, S; Stolzmann, P; Schmid, F T; Scheffel, H; Stinn, B; Marincek, B; Alkadhi, H; Wildermuth, S (2008). Low kilovoltage cardiac dual-source CT: attenuation, noise, and radiation dose. European Radiology, 18(9):1809-1817.

Abstract

The purpose of this study was to investigate the effect of low kilovoltage dual-source computed tomography coronary angiography (CTCA) on qualitative and quantitative image quality parameters and radiation dose. Dual-source CTCA with retrospective ECG gating was performed in 80 consecutive patients of normal weight. Forty were examined with a standard protocol (120 kV/330mAs), 20 were examined at 100 kV/330mAs, and 20 at 100 kV/220mAs. Two blinded observers independently assessed image quality of each coronary segment and measured the image parameters noise, attenuation, and contrast-to-noise ratio (CNR). The effective radiation dose was calculated using CT dose volume index and the dose-length product. Diagnostic image quality was obtained in 99% of all coronary segments (1,127/1,140) without significant differences among the protocols. Image noise, attenuation, and CNR were significantly higher for 100 kV/330mAs (26 +/- 3 HU, 549 +/- 62 HU, 25.5 +/- 3.2; each P < 0.01) and 100 kV/220mAs (27 +/- 2 HU, 560 +/- 43 HU, 25.0 +/- 2.2; each P < 0.01) when compared to the 120-kV protocol (21 +/- 2 HU, 317 +/- 28 HU, 20.6 +/- 1.7). There was no significant difference between the two 100-kV protocols. Estimated effective radiation dose of the 120-kV protocol (8.9 +/- 1.2 mSv) was significantly higher than the 100 kV/330mAs (6.7 +/- 0.8 mSv, P < 0.01) or 100 kV/220mAs (4.4 +/- 0.6 mSv, P < 0.001) protocols. Dual-source CTCA with 100 kV is feasible in patients of normal weight, results in a diagnostic image quality with a higher CNR, and at the same time significantly reduces the radiation dose.

The purpose of this study was to investigate the effect of low kilovoltage dual-source computed tomography coronary angiography (CTCA) on qualitative and quantitative image quality parameters and radiation dose. Dual-source CTCA with retrospective ECG gating was performed in 80 consecutive patients of normal weight. Forty were examined with a standard protocol (120 kV/330mAs), 20 were examined at 100 kV/330mAs, and 20 at 100 kV/220mAs. Two blinded observers independently assessed image quality of each coronary segment and measured the image parameters noise, attenuation, and contrast-to-noise ratio (CNR). The effective radiation dose was calculated using CT dose volume index and the dose-length product. Diagnostic image quality was obtained in 99% of all coronary segments (1,127/1,140) without significant differences among the protocols. Image noise, attenuation, and CNR were significantly higher for 100 kV/330mAs (26 +/- 3 HU, 549 +/- 62 HU, 25.5 +/- 3.2; each P < 0.01) and 100 kV/220mAs (27 +/- 2 HU, 560 +/- 43 HU, 25.0 +/- 2.2; each P < 0.01) when compared to the 120-kV protocol (21 +/- 2 HU, 317 +/- 28 HU, 20.6 +/- 1.7). There was no significant difference between the two 100-kV protocols. Estimated effective radiation dose of the 120-kV protocol (8.9 +/- 1.2 mSv) was significantly higher than the 100 kV/330mAs (6.7 +/- 0.8 mSv, P < 0.01) or 100 kV/220mAs (4.4 +/- 0.6 mSv, P < 0.001) protocols. Dual-source CTCA with 100 kV is feasible in patients of normal weight, results in a diagnostic image quality with a higher CNR, and at the same time significantly reduces the radiation dose.

Citations

163 citations in Web of Science®
190 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

2 downloads since deposited on 29 Sep 2008
0 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Diagnostic and Interventional Radiology
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:2008
Deposited On:29 Sep 2008 10:37
Last Modified:13 Sep 2016 07:29
Publisher:Springer
ISSN:0938-7994
Additional Information:The original publication is available at www.springerlink.com
Publisher DOI:10.1007/s00330-008-0966-1
PubMed ID:18392829
Permanent URL: http://doi.org/10.5167/uzh-3589

Download

[img]
Filetype: PDF - Registered users only
Size: 1MB
View at publisher

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.

Author Collaborations