High-pitch 128-slice dual-source CT for the assessment of coronary stents in a phantom model


Donati, O F; Burg, M C; Desbiolles, L; Karlo, C; Stolzmann, P; Bunck, A; Baumueller, S; Marincek, B; Seifarth, H; Alkadhi, H; Leschka, S; Maintz, D (2010). High-pitch 128-slice dual-source CT for the assessment of coronary stents in a phantom model. Academic Radiology, 17(11):1366-1374.

Abstract

RATIONALE AND OBJECTIVES: To evaluate the quality of stent lumen delineation using dual-source computed tomography (DSCT) in the standard-pitch mode (SP) as compared to the high-pitch mode (HP) in a phantom study.

MATERIALS AND METHODS: Forty different coronary stents placed in plastic tubes filled with contrast agent were imaged with a second generation DSCT system in a SP (pitch 0.23) and HP (pitch 3.4) mode in orientations of 0°, 45°, and 90° relative to the z-axis. Two observers independently measured the in-stent lumen and the attenuation values in the center of the stents. The artificial lumen narrowing (ALN) was calculated using the measured in-stent lumen and the nominal diameter of the plastic tube.

RESULTS: Interobserver correlation was excellent for in-stent lumen (0.86) and attenuation measurements (0.91). There was no significant difference neither for ALN (SP: 54.7-62.8%; HP: 55.8-64.0%) nor attenuation (SP: 356-395 Hounsfield units [HU]; HP: 352-384 HU) between SP and HP mode. For both modes, the orientation of the stent relative to the z-axis significantly affected ALN and attenuation (each P < .001). CT volume dose index was significantly lower using HP mode as compared to SP mode (P < .001).

CONCLUSION: The HP mode in DSCT provides visualization of the coronary in-stent lumen comparable to that measured in SP mode while reducing applied radiation dose in a stationary phantom model.

Abstract

RATIONALE AND OBJECTIVES: To evaluate the quality of stent lumen delineation using dual-source computed tomography (DSCT) in the standard-pitch mode (SP) as compared to the high-pitch mode (HP) in a phantom study.

MATERIALS AND METHODS: Forty different coronary stents placed in plastic tubes filled with contrast agent were imaged with a second generation DSCT system in a SP (pitch 0.23) and HP (pitch 3.4) mode in orientations of 0°, 45°, and 90° relative to the z-axis. Two observers independently measured the in-stent lumen and the attenuation values in the center of the stents. The artificial lumen narrowing (ALN) was calculated using the measured in-stent lumen and the nominal diameter of the plastic tube.

RESULTS: Interobserver correlation was excellent for in-stent lumen (0.86) and attenuation measurements (0.91). There was no significant difference neither for ALN (SP: 54.7-62.8%; HP: 55.8-64.0%) nor attenuation (SP: 356-395 Hounsfield units [HU]; HP: 352-384 HU) between SP and HP mode. For both modes, the orientation of the stent relative to the z-axis significantly affected ALN and attenuation (each P < .001). CT volume dose index was significantly lower using HP mode as compared to SP mode (P < .001).

CONCLUSION: The HP mode in DSCT provides visualization of the coronary in-stent lumen comparable to that measured in SP mode while reducing applied radiation dose in a stationary phantom model.

Citations

5 citations in Web of Science®
8 citations in Scopus®
Google Scholar™

Altmetrics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Diagnostic and Interventional Radiology
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:November 2010
Deposited On:15 Nov 2010 12:01
Last Modified:05 Apr 2016 14:16
Publisher:Elsevier
ISSN:1076-6332
Publisher DOI:https://doi.org/10.1016/j.acra.2010.06.017
PubMed ID:20801697

Download

Full text not available from this repository.
View at publisher

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.

Author Collaborations