Permanent URL to this publication: http://dx.doi.org/10.5167/uzh-38394
Model, D; Eizenman, M; Sturm, V (2010). Fixation-free assessment of the Hirschberg ratio. Investigative Ophthalmology and Visual Science, 51(8):4035-4039.
| PDF - Registered users only 267Kb |
Abstract
PURPOSE: To describe a novel methodology by which to measure the Hirschberg ratio (HR) in infants. The methodology does not require fixation on specific points, and measurements are made while infants look naturally at a display.
METHODS: The HR is calculated automatically from measurements of the direction of the optical axis, the position of the pupil center, and corneal reflexes in video images from an advanced two-camera eye-tracking system. The performance of the novel fixation-free procedure (FFP) was evaluated in 43 adults by measuring the average difference and the 95% limits of agreement with the standard fixation-based procedure (FBP). Repeatability of the HR measurements was evaluated by assessing the 95% limits of agreement between two independent measurements. Performance of the FFP was also evaluated in five infants.
RESULTS: In adults, the average HR was 12.89 +/- 1.22 degrees/mm for FFP and 12.81 +/- 1.22 degrees/mm for FBP. FFP and FBP measurements were highly correlated (r = 0.95; P < 0.001). The 95% limits of agreement between FFP and FBP were +/-0.86 degrees/mm. The 95% limits of agreement of repeated measurements were +/-0.66 degrees/mm for FFP and +/-0.77 degrees/mm for FBP. In infants, the 95% limits of agreement of repeated measurements by FFP were +/-0.63 degrees/mm.
CONCLUSIONS: In adults, the FFP provides accurate measurements of the HR that are in excellent agreement with measurements by FBP. In infants, measurements of HR by FFP show the same repeatability and consistency.
| Item Type: | Journal Article, refereed, original work |
|---|---|
| Communities & Collections: | 04 Faculty of Medicine > University Hospital Zurich > Ophthalmology Clinic |
| DDC: | 610 Medicine & health |
| Language: | English |
| Date: | 2010 |
| Deposited On: | 16 Nov 2010 09:43 |
| Last Modified: | 23 Dec 2012 03:32 |
| Publisher: | Association for Research in Vision and Ophthalmology |
| ISSN: | 0146-0404 |
| PubMed ID: | 20375328 |
| WoS Citation Count: | 3 |