On the reproducibility and repeatability of laser absorption spectroscopy measurements for δ2H and δ18O isotopic analysis


Penna, D; Stenni, B; Šanda, M; Wrede, S; Bogaard, T A; Gobbi, A; Borga, M; Fischer, B M C; Bonazza, M; Chárová, Z (2010). On the reproducibility and repeatability of laser absorption spectroscopy measurements for δ2H and δ18O isotopic analysis. Hydrology and Earth System Sciences, 14(8):1551-1566.

Abstract

The aim of this study was to analyse the reproducibility of off-axis integrated cavity output spectroscopy (OA-ICOS)-derived δ2H and δ18O measurements on a set of 35 water samples by comparing the performance of four laser spectroscopes with the performance of a conventional mass spectrometer under typical laboratory conditions. All samples were analysed using three different schemes of standard/sample combinations and related data processing to assess the improvement of results compared with mass spectrometry. The repeatability of the four OA-ICOS instruments was further investigated by multiple analyses of a sample subset to evaluate the stability of δ2H and δ18O measurements.
Results demonstrated an overall agreement between OA-ICOS-based and mass spectrometry-based measurements for the entire dataset. However, a certain degree of variability existed in precision and accuracy between the four instruments. There was no evident bias or systematic deviations from the mass spectrometer values, but random errors, which were apparently not related to external factors, significantly affected the final results. Our investigation revealed that analytical precision ranged ±from ±0.56‰ to ±1.80‰ for δ2H and from ±0.10‰ to ±0.27‰ for δ18O measurements, with a marked variability among the four instruments. The overall capability of laser instruments to reproduce stable results with repeated measurements of the same sample was acceptable, and there were general differences within the range of the analytical precision for each spectroscope. Hence, averaging the measurements of three identical samples led to a higher degree of accuracy and eliminated the potential for random deviations.

The aim of this study was to analyse the reproducibility of off-axis integrated cavity output spectroscopy (OA-ICOS)-derived δ2H and δ18O measurements on a set of 35 water samples by comparing the performance of four laser spectroscopes with the performance of a conventional mass spectrometer under typical laboratory conditions. All samples were analysed using three different schemes of standard/sample combinations and related data processing to assess the improvement of results compared with mass spectrometry. The repeatability of the four OA-ICOS instruments was further investigated by multiple analyses of a sample subset to evaluate the stability of δ2H and δ18O measurements.
Results demonstrated an overall agreement between OA-ICOS-based and mass spectrometry-based measurements for the entire dataset. However, a certain degree of variability existed in precision and accuracy between the four instruments. There was no evident bias or systematic deviations from the mass spectrometer values, but random errors, which were apparently not related to external factors, significantly affected the final results. Our investigation revealed that analytical precision ranged ±from ±0.56‰ to ±1.80‰ for δ2H and from ±0.10‰ to ±0.27‰ for δ18O measurements, with a marked variability among the four instruments. The overall capability of laser instruments to reproduce stable results with repeated measurements of the same sample was acceptable, and there were general differences within the range of the analytical precision for each spectroscope. Hence, averaging the measurements of three identical samples led to a higher degree of accuracy and eliminated the potential for random deviations.

Citations

28 citations in Web of Science®
33 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

121 downloads since deposited on 29 Dec 2010
31 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:07 Faculty of Science > Institute of Geography
Dewey Decimal Classification:910 Geography & travel
Language:English
Date:2010
Deposited On:29 Dec 2010 16:56
Last Modified:05 Apr 2016 14:26
Publisher:Copernicus
ISSN:1027-5606
Publisher DOI:10.5194/hess-14-1551-2010
Official URL:http://www.hydrol-earth-syst-sci.net/14/1551/2010/
Permanent URL: http://doi.org/10.5167/uzh-39637

Download

[img]
Preview
Filetype: PDF
Size: 3MB
View at publisher

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.

Author Collaborations