Permanent URL to this publication: http://dx.doi.org/10.5167/uzh-41617
Kretschmar, O; Sglimbea, A; Corti, R; Knirsch, W (2010). Shunt reduction with a fenestrated Amplatzer device. Catheterization and Cardiovascular Interventions, 76(4):564-571.
| PDF (Verlags-PDF) - Registered users only 1345Kb | ||
| Accepted Version PDF - Registered users only 2485Kb |
Abstract
BACKGROUND: In specific high-risk patients with congenital heart disease (CHD), a complete closure of an intracardiac defect/shunt is not possible for a variety of reasons. We report our experiences with an interventional approach for shunt-reduction using various modifications of a self-fabricated Amplatzer device in our institution.
METHODS: Retrospective analysis of patients with CHD having received an interventional partial shunt occlusion since 09/2005.
RESULTS: Five patients, mean age 18.6(3.4-66) years, mean weight 36.4(14-102) kg, have been treated. In three patients (3.4, 3.9, 66 years) with an atrial septal defect (ASD) and a restrictive left ventricle (LV) (n = 1) or pulmonary arterial hypertension (PAH) (n = 2), respectively, an Amplatzer Septal Occluder (ASO) with a predilated (n = 2) or a presutured (n = 1) central hole was implanted. After successful immediate volume release in all, the balloon-dilated holes closed spontaneously during mid-term follow-up, pulmonary artery (PA) pressure and LV function remained normal. Two patients (2.7 and 17 years) with a Fontan circulation and severe cyanosis (saturation ≤80%) due to a large fenestration and elevated PA pressures received a partial occlusion of their shunt by implanting a centrally stented ASO or Amplatzer Vascular plug. After a follow-up of 31 and 39 months both stents remained patent under oral anticoagulation, oxygen saturation remained >85% with PA pressures unchanged, and both patients were in good clinical conditions.
CONCLUSIONS: In patients with an ASD and significant PAH and/or restrictive LV physiology as well as in Fontan patients with a large surgically created fenestration but failing Fontan circulation, a partial closure with a self-fenestrated Amplatzer device can be a feasible and successful therapeutic option. Balloon-dilated fenestrations in the Amplatzer device tend to close spontaneously during follow-up. Nonresorbable sutures or stenting can ensure patency of the created holes.
© 2010 Wiley-Liss, Inc.
| Item Type: | Journal Article, refereed, original work |
|---|---|
| Communities & Collections: | 04 Faculty of Medicine > University Children's Hospital Zurich > Medical Clinic 04 Faculty of Medicine > University Hospital Zurich > Clinic for Cardiology |
| DDC: | 610 Medicine & health |
| Language: | English |
| Date: | 2010 |
| Deposited On: | 22 Jan 2011 10:51 |
| Last Modified: | 23 Nov 2012 17:00 |
| Publisher: | Wiley-Blackwell |
| ISSN: | 1522-1946 |
| Publisher DOI: | 10.1002/ccd.22556 |
| PubMed ID: | 20882662 |
Users (please log in): suggest update or correction for this item
Repository Staff Only: item control page