In vitro evaluation of the CeVOX continuous central venous oxygenation monitoring system


Baulig, W; Dullenkopf, A; Hasenclever, P; Schmid, E R; Weiss, M (2008). In vitro evaluation of the CeVOX continuous central venous oxygenation monitoring system. Anaesthesia, 63(4):412-417.

Abstract

We investigated the in vitro performance of the CeVOX system for continuous monitoring of central venous oxygen saturation by spectrophotometry (Pulsion Medical Systems, Munich, Germany). Oxygen inflow into the system was varied, and oxygen saturation values measured by CeVOX were documented. Blood samples were simultaneously taken to assess oxygen saturation by co-oximetry, and values were compared by Bland-Altman analysis. Sixty-six data pairs were obtained at CeVOX and co-oximetry values of 16-99% and 5.5-100%, respectively. Overall, CeVOX values only slightly overestimated co-oximetry values (mean bias +2.4%), but limits of agreement (2 SD of bias) were wide (-11.8 to +16.6%). Saturation measured by CeVOX underestimated that measured by co-oximetry at higher oxygen concentrations and overestimated it at lower oxygen concentrations. There was a nearly linear correlation of the mean bias, suggesting a systematic error. We conclude that the current version of the CeVOX system does not reliably reflect oxygen saturation.

We investigated the in vitro performance of the CeVOX system for continuous monitoring of central venous oxygen saturation by spectrophotometry (Pulsion Medical Systems, Munich, Germany). Oxygen inflow into the system was varied, and oxygen saturation values measured by CeVOX were documented. Blood samples were simultaneously taken to assess oxygen saturation by co-oximetry, and values were compared by Bland-Altman analysis. Sixty-six data pairs were obtained at CeVOX and co-oximetry values of 16-99% and 5.5-100%, respectively. Overall, CeVOX values only slightly overestimated co-oximetry values (mean bias +2.4%), but limits of agreement (2 SD of bias) were wide (-11.8 to +16.6%). Saturation measured by CeVOX underestimated that measured by co-oximetry at higher oxygen concentrations and overestimated it at lower oxygen concentrations. There was a nearly linear correlation of the mean bias, suggesting a systematic error. We conclude that the current version of the CeVOX system does not reliably reflect oxygen saturation.

Citations

5 citations in Web of Science®
5 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

60 downloads since deposited on 03 Nov 2008
14 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Children's Hospital Zurich > Clinic for Surgery
04 Faculty of Medicine > University Hospital Zurich > Institute of Anesthesiology
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:2008
Deposited On:03 Nov 2008 10:00
Last Modified:05 Apr 2016 12:32
Publisher:Wiley-Blackwell
ISSN:0003-2409
Additional Information:The definitive version is available at www.blackwell-synergy.com
Publisher DOI:10.1111/j.1365-2044.2007.05376.x
PubMed ID:18336492
Permanent URL: http://doi.org/10.5167/uzh-4840

Download

[img]
Preview
Content: Accepted Version
Filetype: PDF
Size: 1MB
View at publisher

[img]Filetype: PDF - Registered users only
Size: 1MB

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.

Author Collaborations