Baulig, W; Dullenkopf, A; Hasenclever, P; Schmid, E R; Weiss, M (2008). In vitro evaluation of the CeVOX continuous central venous oxygenation monitoring system. Anaesthesia, 63(4):412-417.
| Accepted Version 1264Kb | |
| PDF - Registered users only 1198Kb |
Abstract
We investigated the in vitro performance of the CeVOX system for continuous monitoring of central venous oxygen saturation by spectrophotometry (Pulsion Medical Systems, Munich, Germany). Oxygen inflow into the system was varied, and oxygen saturation values measured by CeVOX were documented. Blood samples were simultaneously taken to assess oxygen saturation by co-oximetry, and values were compared by Bland-Altman analysis. Sixty-six data pairs were obtained at CeVOX and co-oximetry values of 16-99% and 5.5-100%, respectively. Overall, CeVOX values only slightly overestimated co-oximetry values (mean bias +2.4%), but limits of agreement (2 SD of bias) were wide (-11.8 to +16.6%). Saturation measured by CeVOX underestimated that measured by co-oximetry at higher oxygen concentrations and overestimated it at lower oxygen concentrations. There was a nearly linear correlation of the mean bias, suggesting a systematic error. We conclude that the current version of the CeVOX system does not reliably reflect oxygen saturation.
| Item Type: | Journal Article, refereed, original work |
|---|---|
| Communities & Collections: | 04 Faculty of Medicine > University Children's Hospital Zurich > Clinic for Surgery 04 Faculty of Medicine > University Hospital Zurich > Institute of Anesthesiology |
| DDC: | 610 Medicine & health |
| Language: | English |
| Date: | 2008 |
| Deposited On: | 03 Nov 2008 11:00 |
| Last Modified: | 23 Nov 2012 13:39 |
| Publisher: | Wiley-Blackwell |
| ISSN: | 0003-2409 |
| Additional Information: | The definitive version is available at www.blackwell-synergy.com |
| Publisher DOI: | 10.1111/j.1365-2044.2007.05376.x |
| PubMed ID: | 18336492 |
Users (please log in): suggest update or correction for this item
Repository Staff Only: item control page