UZH-Logo

Maintenance Infos

Use of a single bipolar electrode in the posterior arytenoid muscles for bilateral monitoring of the recurrent laryngeal nerves in thyroid surgery


Haerle, S; Sidler, D; Linder, T; Mueller, W (2008). Use of a single bipolar electrode in the posterior arytenoid muscles for bilateral monitoring of the recurrent laryngeal nerves in thyroid surgery. European Archives of Oto-Rhino-Laryngology, 265(12):1549-1552.

Abstract

The aims were to assess the technical feasibility of using a single electrode in the posterior arytenoid muscles (PAM) for intraoperative monitoring of the recurrent laryngeal nerve (RLN) in thyroid surgery, to validate the new method against the insertion of electrodes placed in the vocal cord muscle, and to report the results of the clinical application of the new concept. A total of 52 patients were enrolled. The handling and safety of RLN monitoring was tested by simultaneous registration of the EMG response from vocal fold electrodes and PAM electrodes. Acoustically and electromyographically we found nearly the same values for the arytenoid muscles as for the vocal folds, although the signals taken from the vocal folds were slightly stronger. PAM recording using a single bipolar electrode is technically feasible and as reliable compared to the standard vocal cord monitoring.

The aims were to assess the technical feasibility of using a single electrode in the posterior arytenoid muscles (PAM) for intraoperative monitoring of the recurrent laryngeal nerve (RLN) in thyroid surgery, to validate the new method against the insertion of electrodes placed in the vocal cord muscle, and to report the results of the clinical application of the new concept. A total of 52 patients were enrolled. The handling and safety of RLN monitoring was tested by simultaneous registration of the EMG response from vocal fold electrodes and PAM electrodes. Acoustically and electromyographically we found nearly the same values for the arytenoid muscles as for the vocal folds, although the signals taken from the vocal folds were slightly stronger. PAM recording using a single bipolar electrode is technically feasible and as reliable compared to the standard vocal cord monitoring.

Citations

3 citations in Web of Science®
2 citations in Scopus®
Google Scholar™

Altmetrics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Otorhinolaryngology
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:2008
Deposited On:04 Nov 2008 16:52
Last Modified:05 Apr 2016 12:32
Publisher:Springer
ISSN:0937-4477
Publisher DOI:10.1007/s00405-008-0671-3
PubMed ID:18415115

Download

Full text not available from this repository.View at publisher

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.