Quick Search:

uzh logo
Browse by:
bullet
bullet
bullet
bullet

Zurich Open Repository and Archive

Nordmeyer-Massner, J A; Wyss, M; Andreisek, G; Pruessmann, K P; Hodler, J (2011). In vitro and in vivo comparison of wrist MR imaging at 3.0 and 7.0 tesla using a gradient echo sequence and identical eight-channel coil array designs. Journal of Magnetic Resonance Imaging, 33(3):661-667.

Full text not available from this repository.

View at publisher

Abstract

The presented results establish the SNR benefits of the transition from 3.0T to 7.0T for wrist imaging without bias by different array designs and based on exact, algebraic SNR quantification. The observed SNR increase nearly reaches expected values but varies greatly between different tissues. It does not necessarily improve the visibility of anatomic structures but adds valuable latitude for sequence optimization.

Citations

8 citations in Web of Science®
8 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

0 downloads since deposited on 21 Nov 2011
0 downloads since 12 months

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Diagnostic and Interventional Radiology
04 Faculty of Medicine > Institute of Biomedical Engineering
DDC:170 Ethics
610 Medicine & health
Language:English
Date:2011
Deposited On:21 Nov 2011 13:14
Last Modified:27 Nov 2013 22:13
Publisher:Wiley-Blackwell
ISSN:1053-1807
Publisher DOI:10.1002/jmri.22419
PubMed ID:21563250

Users (please log in): suggest update or correction for this item

Repository Staff Only: item control page