Sustained growth of the ex vivo ablation zones' critical short axis using gas-cooled radiofrequency applicators


Rempp, H; Scharpf, M; Voigtlaender, M; Schraml, C; Schmidt, D; Fend, F; Claussen, C D; Enderle, M D; Pereira, P L; Clasen, S (2011). Sustained growth of the ex vivo ablation zones' critical short axis using gas-cooled radiofrequency applicators. Cardiovascular and Interventional Radiology, 34(1):149-155.

Abstract

Purpose
To evaluate the ablation zones created with a gas-cooled bipolar radiofrequency applicator performed on ex vivo bovine liver tissue.
Materials and Methods
A total of 320 ablations with an internally gas-cooled bipolar radiofrequency applicator were performed on fresh ex vivo bovine liver tissue, varying the ablation time (5, 10, 15, and 20 min), power (20, 30, 40, and 50 W), and gas pressure of the CO2 used for cooling (585, 600, 615, 630, 645 psi), leading to a total of 80 different parameter combinations. Size and shape of the white coagulation zone were assessed.
Results
The largest complete ablation zone was achieved after 20 min of implementing 50 W and 645 psi, resulting in a short axis of mean 46 ± 1 mm and a long axis of 56 ± 2 mm (mean ± standard deviation). Short-axis diameters increased between 5 and 20 min of ablation time at 585 psi (increase of the short axis was 45% at 30 W, 29% at 40 W, and 39% at 50 W). This increase was larger at 645 psi (113% at 30 W, 67% at 40 W, and 70% at 50 W). Macroscopic assessment and NADH (nicotinamide adenine dinucleotide) staining revealed incompletely ablated tissue along the needle track in 18 parameter combinations including low-power settings (20 and 30 W) and different cooling levels and ablation times.
Conclusion
Gas-cooled radiofrequency applicators increase the short-axis diameter of coagulation in an ex vivo setting if appropriate parameters are selected.

Abstract

Purpose
To evaluate the ablation zones created with a gas-cooled bipolar radiofrequency applicator performed on ex vivo bovine liver tissue.
Materials and Methods
A total of 320 ablations with an internally gas-cooled bipolar radiofrequency applicator were performed on fresh ex vivo bovine liver tissue, varying the ablation time (5, 10, 15, and 20 min), power (20, 30, 40, and 50 W), and gas pressure of the CO2 used for cooling (585, 600, 615, 630, 645 psi), leading to a total of 80 different parameter combinations. Size and shape of the white coagulation zone were assessed.
Results
The largest complete ablation zone was achieved after 20 min of implementing 50 W and 645 psi, resulting in a short axis of mean 46 ± 1 mm and a long axis of 56 ± 2 mm (mean ± standard deviation). Short-axis diameters increased between 5 and 20 min of ablation time at 585 psi (increase of the short axis was 45% at 30 W, 29% at 40 W, and 39% at 50 W). This increase was larger at 645 psi (113% at 30 W, 67% at 40 W, and 70% at 50 W). Macroscopic assessment and NADH (nicotinamide adenine dinucleotide) staining revealed incompletely ablated tissue along the needle track in 18 parameter combinations including low-power settings (20 and 30 W) and different cooling levels and ablation times.
Conclusion
Gas-cooled radiofrequency applicators increase the short-axis diameter of coagulation in an ex vivo setting if appropriate parameters are selected.

Citations

4 citations in Web of Science®
3 citations in Scopus®
Google Scholar™

Altmetrics

Additional indexing

Item Type:Journal Article, not refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Diagnostic and Interventional Radiology
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:2011
Deposited On:07 Dec 2011 10:16
Last Modified:05 Apr 2016 15:08
Publisher:Springer
ISSN:0174-1551
Publisher DOI:https://doi.org/10.1007/s00270-010-9946-3
PubMed ID:20661567

Download

Full text not available from this repository.
View at publisher

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.