Ulmer, F F; Baulig, W; Béttex, D; Spielmann, N; Bürki, C; Weiss, M (2011). Measurement of activated coagulation time in children: evaluation of the blood-saving kaolin i-STAT activated coagulation time technique in pediatric cardiac anesthesia. Journal of Cardiothoracic and Vascular Anesthesia, 25(3):395-401.
Full text not available from this repository.
Abstract
The kaolin i-STAT 1 ACT technique agreed well with the Medtronic ACTR II technique during the nonheparinized phase that preceded CPB. The overall agreement between the ACT obtained from the 2 devices was poor. The routine use of i-STAT 1 measured ACT values cannot be recommended as a reliable alternative to the Medtronic ACTR II.