Cardiac output measurement in children: comparison of the Ultrasound Cardiac Output Monitor with thermodilution cardiac output measurement


Knirsch, W; Kretschmar, O; Tomaske, M; Stutz, K; Nagdyman, N; Balmer, C; Schmitz, A; Bettex, D; Berger, F; Bauersfeld, U; Weiss, M (2008). Cardiac output measurement in children: comparison of the Ultrasound Cardiac Output Monitor with thermodilution cardiac output measurement. Intensive Care Medicine, 34(6):1060-1064.

Abstract

OBJECTIVE: To compare the assessment of cardiac output (CO) in children using the noninvasive Ultrasound Cardiac Output Monitor (USCOM) with the invasive pulmonary artery catheter (PAC) thermodilution cardiac output measurement. DESIGN AND SETTING: Prospective observational study in a tertiary center for pediatric cardiology of a university children's hospital. PATIENTS: Twenty-four pediatric patients with congenital heart disease without shunt undergoing cardiac catheterization under general anesthesia. MEASUREMENTS AND RESULTS: CO was measured by USCOM using a suprasternal CO Doppler probe in children undergoing cardiac catheterization. USCOM data were compared to CO simultaneously measured by PAC thermodilution technique. Measurements were repeated three times within 5 min in each patient. A mean percentage error not exceeding 30% was defined as indicating clinical useful reliability of the USCOM. CO values measured by PAC ranged from 1.3 to 5.3 l/min (median 3.6 l/min). Bias and precision were -0.13 and 1.34 l/min, respectively. The mean percentage error of CO measurement by the USCOM compared to PAC thermodilution technique was 36.4% for USCOM. CONCLUSIONS: Our preliminary data demonstrate that cardiac output measurement in children using the USCOM does not reliably represent absolute CO values as compared to PAC thermodilution. Further studies must evaluate the impact of incorporating effective aortic valve diameters on CO measurement using the USCOM.

OBJECTIVE: To compare the assessment of cardiac output (CO) in children using the noninvasive Ultrasound Cardiac Output Monitor (USCOM) with the invasive pulmonary artery catheter (PAC) thermodilution cardiac output measurement. DESIGN AND SETTING: Prospective observational study in a tertiary center for pediatric cardiology of a university children's hospital. PATIENTS: Twenty-four pediatric patients with congenital heart disease without shunt undergoing cardiac catheterization under general anesthesia. MEASUREMENTS AND RESULTS: CO was measured by USCOM using a suprasternal CO Doppler probe in children undergoing cardiac catheterization. USCOM data were compared to CO simultaneously measured by PAC thermodilution technique. Measurements were repeated three times within 5 min in each patient. A mean percentage error not exceeding 30% was defined as indicating clinical useful reliability of the USCOM. CO values measured by PAC ranged from 1.3 to 5.3 l/min (median 3.6 l/min). Bias and precision were -0.13 and 1.34 l/min, respectively. The mean percentage error of CO measurement by the USCOM compared to PAC thermodilution technique was 36.4% for USCOM. CONCLUSIONS: Our preliminary data demonstrate that cardiac output measurement in children using the USCOM does not reliably represent absolute CO values as compared to PAC thermodilution. Further studies must evaluate the impact of incorporating effective aortic valve diameters on CO measurement using the USCOM.

Citations

27 citations in Web of Science®
32 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

1 download since deposited on 18 Nov 2008
0 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Children's Hospital Zurich > Clinic for Surgery
04 Faculty of Medicine > University Hospital Zurich > Institute of Anesthesiology
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:2008
Deposited On:18 Nov 2008 16:28
Last Modified:05 Apr 2016 12:33
Publisher:Springer
ISSN:0342-4642
Publisher DOI:10.1007/s00134-008-1030-y
PubMed ID:18297271
Permanent URL: http://doi.org/10.5167/uzh-5351

Download

[img]
Filetype: PDF - Registered users only
Size: 1MB
View at publisher

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.

Author Collaborations