In vitro assessments of reverse glenoid stability using displacement gages are misleading - recommendations for accurate measurements of interface micromotion


Favre, P; Perala, S; Vogel, P; Fucentese, S F; Goff, J R; Gerber, C; Snedeker, J G (2011). In vitro assessments of reverse glenoid stability using displacement gages are misleading - recommendations for accurate measurements of interface micromotion. Clinical Biomechanics, 26(9):917-922.

Abstract

BACKGROUND:

Baseplate micromotion of the reverse shoulder glenoid component can lead to implant loosening. We hypothesized that a remotely positioned displacement gage measures elastic deformation of the system rather than actual micromotion at the implant/bone interface.
METHODS:

Reverse glenoid components were implanted into polyurethane blocks of 3 different densities. A 700 N compressive load was maintained and a vertical 700 N shear load was applied for 1000 cycles. In addition to the typical gage measurement, a digital image analysis of micromotion at the implant/block interface using high resolution cameras was performed. The measurements were validated on human specimens. A finite element model was implemented to study the isolated effect of block deformation on baseplate displacements.
FINDINGS:

With the gage, typically reported micromotions were measured. Two orders of magnitude lower micromotions were detected using interface image-based analysis. The finite element simulation showed that elastic deformation alone can cause micromotion magnitudes as measured with displacement gages. Polyurethane blocks of 20 and 15 lbs per cubic foot density best reproduced micromotions as measured on human specimens.
INTERPRETATION:

We found considerably less relative micromotion at the implant/bone interface than previously assumed. Gage measurements quantify elastic deformation and not true interface micromotion. High resolution digital imaging at the implant/bone interface is strongly recommended for an accurate assessment of reverse glenoid component micromotion. Tests should further adopt 20 or 15 pcf bone test surrogates. Further studies are required to identify the failure modes encountered in vivo, and a corresponding in vitro testing methodology can then be developed.

Copyright © 2011 Elsevier Ltd. All rights reserved.

BACKGROUND:

Baseplate micromotion of the reverse shoulder glenoid component can lead to implant loosening. We hypothesized that a remotely positioned displacement gage measures elastic deformation of the system rather than actual micromotion at the implant/bone interface.
METHODS:

Reverse glenoid components were implanted into polyurethane blocks of 3 different densities. A 700 N compressive load was maintained and a vertical 700 N shear load was applied for 1000 cycles. In addition to the typical gage measurement, a digital image analysis of micromotion at the implant/block interface using high resolution cameras was performed. The measurements were validated on human specimens. A finite element model was implemented to study the isolated effect of block deformation on baseplate displacements.
FINDINGS:

With the gage, typically reported micromotions were measured. Two orders of magnitude lower micromotions were detected using interface image-based analysis. The finite element simulation showed that elastic deformation alone can cause micromotion magnitudes as measured with displacement gages. Polyurethane blocks of 20 and 15 lbs per cubic foot density best reproduced micromotions as measured on human specimens.
INTERPRETATION:

We found considerably less relative micromotion at the implant/bone interface than previously assumed. Gage measurements quantify elastic deformation and not true interface micromotion. High resolution digital imaging at the implant/bone interface is strongly recommended for an accurate assessment of reverse glenoid component micromotion. Tests should further adopt 20 or 15 pcf bone test surrogates. Further studies are required to identify the failure modes encountered in vivo, and a corresponding in vitro testing methodology can then be developed.

Copyright © 2011 Elsevier Ltd. All rights reserved.

Citations

5 citations in Web of Science®
4 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

8292 downloads since deposited on 07 Feb 2012
8170 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > Balgrist University Hospital, Swiss Spinal Cord Injury Center
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:2011
Deposited On:07 Feb 2012 09:48
Last Modified:05 Apr 2016 15:32
Publisher:Elsevier
ISSN:0268-0033 (P) 1879-1271 (E)
Publisher DOI:https://doi.org/10.1016/j.clinbiomech.2011.05.002
PubMed ID:21658824
Permanent URL: https://doi.org/10.5167/uzh-57885

Download

[img]
Preview
Content: Accepted Version
Language: English
Filetype: PDF
Size: 3MB
View at publisher

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.

Author Collaborations