Assessment of a direct acoustic cochlear stimulator


Chatzimichalis, M; Sim, J H; Huber, A M (2012). Assessment of a direct acoustic cochlear stimulator. Audiology & Neurotology, 17(5):299-308.

Abstract

This study aimed to assess the functional results of a new, active, acoustic-mechanical hearing implant, the Direct Acoustic Cochlear Stimulation Partial Implant (DACS PI), in a preclinical study. The DACS PI is an electromagnetic device fixed to the mastoid by screws and coupled to a standard stapes prosthesis by an artificial incus (AI). The function of the DACS PI-aided reconstruction was assessed by determining: (1) the maximum equivalent sound pressure level (SPL) of the implant, which was obtained from measurements of the volume displacement at the round window in normal and implanted ears, and (2) the quality at the coupling interface between the AI of the DACS and the stapes prosthesis, which was quantified from measurements of relative motions between the AI and the prosthesis. Both measurements were performed with fresh temporal bones using a scanning laser Doppler interferometry system. The expected maximum equivalent SPL with a typical driving voltage of 0.3 V was about 115-125 dB SPL up to 1.5 kHz in reconstruction with the DACS PI, and decreased with a roll-off slope of about 65 dB/decade, reaching 90 dB SPL at 8 kHz. The large roll-off relative to a normal ear was presumed to be a relatively high inductive impedance of the coil of the DACS PI actuator at higher frequencies. Good coupling quality between the AI and the prosthesis was achieved below the resonance (∼1.5 kHz) of the DACS PI for all tested stapes prostheses. Above the resonance, the SMart Piston, which is composed of a shape-memory alloy, had the best coupling quality.

This study aimed to assess the functional results of a new, active, acoustic-mechanical hearing implant, the Direct Acoustic Cochlear Stimulation Partial Implant (DACS PI), in a preclinical study. The DACS PI is an electromagnetic device fixed to the mastoid by screws and coupled to a standard stapes prosthesis by an artificial incus (AI). The function of the DACS PI-aided reconstruction was assessed by determining: (1) the maximum equivalent sound pressure level (SPL) of the implant, which was obtained from measurements of the volume displacement at the round window in normal and implanted ears, and (2) the quality at the coupling interface between the AI of the DACS and the stapes prosthesis, which was quantified from measurements of relative motions between the AI and the prosthesis. Both measurements were performed with fresh temporal bones using a scanning laser Doppler interferometry system. The expected maximum equivalent SPL with a typical driving voltage of 0.3 V was about 115-125 dB SPL up to 1.5 kHz in reconstruction with the DACS PI, and decreased with a roll-off slope of about 65 dB/decade, reaching 90 dB SPL at 8 kHz. The large roll-off relative to a normal ear was presumed to be a relatively high inductive impedance of the coil of the DACS PI actuator at higher frequencies. Good coupling quality between the AI and the prosthesis was achieved below the resonance (∼1.5 kHz) of the DACS PI for all tested stapes prostheses. Above the resonance, the SMart Piston, which is composed of a shape-memory alloy, had the best coupling quality.

Citations

3 citations in Web of Science®
5 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

1 download since deposited on 18 Sep 2012
0 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Otorhinolaryngology
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:2012
Deposited On:18 Sep 2012 12:02
Last Modified:15 Jun 2016 15:59
Publisher:Karger
ISSN:1421-9700
Publisher DOI:10.1159/000339214
PubMed ID:22739432
Permanent URL: http://doi.org/10.5167/uzh-64708

Download

[img]
Preview
Content: Published Version
Language: English
Filetype: PDF
Size: 1MB
View at publisher

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.

Author Collaborations