Bending moments and types of failure of zirconia and titanium abutments with internal implant-abutment connections: a laboratory study


Leutert, Christian R; Stawarczyk, Bogna; Truninger, Thomas C; Hämmerle, Christoph H F; Sailer, Irena (2012). Bending moments and types of failure of zirconia and titanium abutments with internal implant-abutment connections: a laboratory study. International Journal of Oral and Maxillofacial Implants, 27(3):505-512.

Abstract

PURPOSE: The aim of this study was to examine the bending moments and fracture patterns of different zirconia abutments with internal implant-abutment connections after static loading and to compare their bending moments to those of internally connected titanium abutments. Materials and METHODS: Three types of customized zirconia abutments (Straumann CARES abutments/Straumann BL implants [T1], Astra ZirDesign abutments/Astra Micro Thread OsseoSpeed implants [T2], Zirabut prototype abutments/Straumann SP implants [T3]) and one type of customized titanium abutment (control group, Straumann CARES abutments/Straumann BL implants [C]) were included. All abutments were one-piece abutments with an internal implant-abutment connection and were customized to the same shape but featured different implant-abutment connection designs. For each group, 20 identical copies of a master abutment were fabricated and fixed on their corresponding implants. Half of the abutments in each group were left unrestored, and the other 10 received glass-ceramic crowns. Static loading was applied at a 30-degree angle to the palatal surface until failure, and bending moments were calculated. The type of failure was characterized visually by dismounting the abutments and by examination of cross-sections of the embedded specimens. The results were analyzed statistically.
RESULTS: The mean range of bending moments was higher for the unrestored groups (158.2 to 678.2 Ncm) than for the restored groups (117.9 to 419.4 Ncm). The highest mean bending moments were seen in the control group, both restored and unrestored (419.4/678.2 Ncm). Unrestored, T1 and T2 exhibited significantly higher bending moments than T3. This was also observed in the restored groups. CONCLUSION:Both the abutment material and the implant-abutment connection design affected the bending moments of abutments after static loading. Internally connected zirconia abutments with horizontal mismatch to the implant exhibited significantly higher bending moments compared to those without horizontal mismatch.

Abstract

PURPOSE: The aim of this study was to examine the bending moments and fracture patterns of different zirconia abutments with internal implant-abutment connections after static loading and to compare their bending moments to those of internally connected titanium abutments. Materials and METHODS: Three types of customized zirconia abutments (Straumann CARES abutments/Straumann BL implants [T1], Astra ZirDesign abutments/Astra Micro Thread OsseoSpeed implants [T2], Zirabut prototype abutments/Straumann SP implants [T3]) and one type of customized titanium abutment (control group, Straumann CARES abutments/Straumann BL implants [C]) were included. All abutments were one-piece abutments with an internal implant-abutment connection and were customized to the same shape but featured different implant-abutment connection designs. For each group, 20 identical copies of a master abutment were fabricated and fixed on their corresponding implants. Half of the abutments in each group were left unrestored, and the other 10 received glass-ceramic crowns. Static loading was applied at a 30-degree angle to the palatal surface until failure, and bending moments were calculated. The type of failure was characterized visually by dismounting the abutments and by examination of cross-sections of the embedded specimens. The results were analyzed statistically.
RESULTS: The mean range of bending moments was higher for the unrestored groups (158.2 to 678.2 Ncm) than for the restored groups (117.9 to 419.4 Ncm). The highest mean bending moments were seen in the control group, both restored and unrestored (419.4/678.2 Ncm). Unrestored, T1 and T2 exhibited significantly higher bending moments than T3. This was also observed in the restored groups. CONCLUSION:Both the abutment material and the implant-abutment connection design affected the bending moments of abutments after static loading. Internally connected zirconia abutments with horizontal mismatch to the implant exhibited significantly higher bending moments compared to those without horizontal mismatch.

Citations

21 citations in Web of Science®
9 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

4 downloads since deposited on 02 Oct 2012
0 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > Center for Dental Medicine > Clinic for Fixed and Removable Prosthodontics
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:2012
Deposited On:02 Oct 2012 14:32
Last Modified:05 Apr 2016 15:58
Publisher:Quintessence Publishing
ISSN:0882-2786
Related URLs:http://www.quintpub.com/index.php (Publisher)
PubMed ID:22616042

Download

[img]
Content: Accepted Version
Filetype: PDF - Registered users only
Size: 15MB
[img]
Content: Published Version
Filetype: PDF - Registered users only
Size: 324kB

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.

Author Collaborations