Madjdpour, C; Mauch, J; Dave, M H; Spielmann, N; Weiss, M (2012). Comparison of air-sealing characteristics of tapered- vs. cylindrical-shaped high-volume, low-pressure tube cuffs. Acta Anaesthesiologica Scandinavica, 56(2):230-235.

Abstract

BACKGROUND: This study aimed at comparing air-sealing characteristics of the new tapered-shaped tracheal tube cuffs with cylindrical tube cuffs.
METHODS: Tracheal tubes with tapered-shaped polyurethane (PU) and polyvinyl chloride (PVC) cuffs as well as standard cylindrical-shaped cuffs made of PU and PVC (Covidien, Athlone, Ireland) were investigated. A tracheal model attached to a test lung was intubated, and cuffs were inflated to 10, 15, 20, 25 and 30 cm H(2)O. The test lung was ventilated with intermittent positive pressure ventilation at peak inspiratory pressures (PIPs) of 20 and 25 cm H(2)O. Air leakage was assessed by spirometry and measurement of sevoflurane concentration passing the cuff at the upper cuff border using an anaesthetic gas analyser. Experiments were repeated four times with new tracheal tubes for each run. Statistical comparisons were done using Mann-Whitney U-test with level of significance at P < 0.05.
RESULTS: The tapered-shaped PVC tube cuff demonstrated a significantly lower air leakage determined by spirometry than the cylindrical-shaped cuff at both PIPs (20 and 25 cm H(2)O). Similarly, sevoflurane leakage was less with the tapered PVC cuff particularly at higher cuff pressures. With the PU cuff, reduction in air leakage by a tapered-shaped compared with a cylindrical-shaped tube cuff was not significant.
CONCLUSIONS: A tapered-shaped tube cuff considerably improves air-sealing characteristics of PVC tube cuffs and allows thereby reducing cuff pressure required for sufficient ventilation. In tube cuffs made of PU that exhibits superior sealing characteristics compared with PVC, a tapered shape failed to result in a further reduction of air leakage.

Abstract

BACKGROUND: This study aimed at comparing air-sealing characteristics of the new tapered-shaped tracheal tube cuffs with cylindrical tube cuffs.
METHODS: Tracheal tubes with tapered-shaped polyurethane (PU) and polyvinyl chloride (PVC) cuffs as well as standard cylindrical-shaped cuffs made of PU and PVC (Covidien, Athlone, Ireland) were investigated. A tracheal model attached to a test lung was intubated, and cuffs were inflated to 10, 15, 20, 25 and 30 cm H(2)O. The test lung was ventilated with intermittent positive pressure ventilation at peak inspiratory pressures (PIPs) of 20 and 25 cm H(2)O. Air leakage was assessed by spirometry and measurement of sevoflurane concentration passing the cuff at the upper cuff border using an anaesthetic gas analyser. Experiments were repeated four times with new tracheal tubes for each run. Statistical comparisons were done using Mann-Whitney U-test with level of significance at P < 0.05.
RESULTS: The tapered-shaped PVC tube cuff demonstrated a significantly lower air leakage determined by spirometry than the cylindrical-shaped cuff at both PIPs (20 and 25 cm H(2)O). Similarly, sevoflurane leakage was less with the tapered PVC cuff particularly at higher cuff pressures. With the PU cuff, reduction in air leakage by a tapered-shaped compared with a cylindrical-shaped tube cuff was not significant.
CONCLUSIONS: A tapered-shaped tube cuff considerably improves air-sealing characteristics of PVC tube cuffs and allows thereby reducing cuff pressure required for sufficient ventilation. In tube cuffs made of PU that exhibits superior sealing characteristics compared with PVC, a tapered shape failed to result in a further reduction of air leakage.

Citations

8 citations in Web of Science®
8 citations in Scopus®
Google Scholar™

Altmetrics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Institute of Anesthesiology
04 Faculty of Medicine > University Children's Hospital Zurich > Clinic for Surgery
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:2012
Deposited On:09 Nov 2012 11:27
Last Modified:05 Apr 2016 16:03
Publisher:Wiley-Blackwell
ISSN:0001-5172
Publisher DOI:https://doi.org/10.1111/j.1399-6576.2011.02542.x
PubMed ID:22091784

Download

Full text not available from this repository.
View at publisher

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.

Author Collaborations