Real-time multipoint gastrointestinal 19-fluorine catheter tracking


Hahn, Tobias; Kozerke, Sebastian; Schwizer, Werner; Fried, Michael; Boesiger, Peter; Steingoetter, Andreas (2014). Real-time multipoint gastrointestinal 19-fluorine catheter tracking. Magnetic Resonance in Medicine, 71(1):302-307.

Abstract

PURPOSE: To develop MR based real-time gastrointestinal 19-Fluorine (19F) catheter tracking and visualization allowing for real-time detection and feedback of 3D catheter shape and movement as well as catheter-driven adjustments of 1H imaging geometry parameters.
METHODS: Data were acquired on a 3T clinical system using 3D Golden Angle radial sampling. Two gastrointestinal catheters incorporating four fiducial 19F markers (65 or 50 µL marker volume) were tracked while being pulled through a gel phantom by an operator inside the MR room with velocities of 2-18 mm/s. During continuous acquisition, k-space profiles were transferred in real-time to an external computer for concurrent reconstruction of 3D 19F images and detection and visualization of marker positions. Based on αthe marker positions, automatic adjustments of 1H imaging planes to facilitate targeted anatomical scanning was implemented.
RESULTS: Mean tracking reliabilities were 94.5 and 83.6% (catheters 1 and 2) for temporal resolutions 185-740 ms. Reconstruction times of 196 ms were achieved. Real-time visual feedback allowed the operator to accurately control the catheter movement. Catheter-guidance for 1H imaging was reliable.
CONCLUSION:The presented real-time 19F MR based framework for the tracking of 19F labeled devices is applicable to combined 19F and 1H MRI guidance of gastrointestinal devices in vivo.

Abstract

PURPOSE: To develop MR based real-time gastrointestinal 19-Fluorine (19F) catheter tracking and visualization allowing for real-time detection and feedback of 3D catheter shape and movement as well as catheter-driven adjustments of 1H imaging geometry parameters.
METHODS: Data were acquired on a 3T clinical system using 3D Golden Angle radial sampling. Two gastrointestinal catheters incorporating four fiducial 19F markers (65 or 50 µL marker volume) were tracked while being pulled through a gel phantom by an operator inside the MR room with velocities of 2-18 mm/s. During continuous acquisition, k-space profiles were transferred in real-time to an external computer for concurrent reconstruction of 3D 19F images and detection and visualization of marker positions. Based on αthe marker positions, automatic adjustments of 1H imaging planes to facilitate targeted anatomical scanning was implemented.
RESULTS: Mean tracking reliabilities were 94.5 and 83.6% (catheters 1 and 2) for temporal resolutions 185-740 ms. Reconstruction times of 196 ms were achieved. Real-time visual feedback allowed the operator to accurately control the catheter movement. Catheter-guidance for 1H imaging was reliable.
CONCLUSION:The presented real-time 19F MR based framework for the tracking of 19F labeled devices is applicable to combined 19F and 1H MRI guidance of gastrointestinal devices in vivo.

Altmetrics

Downloads

1 download since deposited on 21 Mar 2013
0 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Gastroenterology and Hepatology
04 Faculty of Medicine > Institute of Biomedical Engineering
Dewey Decimal Classification:170 Ethics
610 Medicine & health
Language:English
Date:2014
Deposited On:21 Mar 2013 15:50
Last Modified:05 Apr 2016 16:34
Publisher:Wiley-Blackwell
ISSN:0740-3194
Publisher DOI:https://doi.org/10.1002/mrm.24654
PubMed ID:23400935

Download

[img]
Filetype: PDF - Registered users only
Size: 1MB
View at publisher

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.

Author Collaborations