TOPOS: a new topometric patient positioning and tracking system for radiation therapy based on structured white light


Lindl, Bastian L; Müller, Reinhold G; Lang, Stephanie; Herraiz Lablanca, Maria D; Klöck, Stephan (2013). TOPOS: a new topometric patient positioning and tracking system for radiation therapy based on structured white light. Medical Physics, 40(4):042701.

Abstract

PURPOSE: A patient positioning system for radiation therapy based on structured white light and using off-the-shelf hardware components for flexibility and cost-effectiveness has been developed in house. Increased accuracy, patient comfort, abandonment of any skin marks, accelerated workflow, objective reading/recording, better usability and robust sensor design, compared to other positioning approaches, were the main goals of this work. Another aim was the application of a 6 degrees of freedom tracking system working without dose deposition. METHODS: Two optical sensors are the main parts of the TOPOS® system (Topometrical Positioning, cyberTECHNOLOGIES, Germany). The components: cameras, projectors, and computers are commercial off-the-shelf products, allowing for low production costs. The black/white cameras of the prototype are capable of taking up to 240 frames per second (resolution: 640 × 488 pixels). The projector has a resolution of 1024 × 768 and a refresh rate of 120 Hz. The patient's body surface is measured continuously and registered to a reference surface, providing a transformation to superimpose the patient's surface to the reference (planning CT) surface as best as possible. The execution of the calculated transformation provides the correct patient position before the treatment starts. Due to the high-speed acquisition of the surfaces, a surveillance of the patient's (respiration) motion during treatment is also accomplished. The accuracy of the system was determined using a male mannequin. Two treatment sites were evaluated: one simulating a head and neck treatment and the other simulating a thoracic wall treatment. The mannequin was moved to predefined positions, and shift vectors given by the surface registration were evaluated. Additionally manual positioning using a color-coding system was evaluated. RESULTS: Two prototypes have been developed, each allowing a continuous high density scan of a 500 × 500 × 400 mm(3) (L × W × D) large volume with a refresh rate of 10 Hz (extendible to 20 Hz for a single sensor system). Surface and position correction display, as well as respiratory motion, is shown in real-time (delay < 200 ms) using present graphical hardware acceleration. For an intuitive view of the patient's misalignment, a fast surface registration algorithm has been developed and tested and a real-time color-coding technique is proposed and verified that allows the user to easily verify the position of the patient. Using first the surface registration and then the color coding the best results were obtained: for the head and neck case, the mean difference between the actual zero position and the final match was 0.1 ± 0.4, -0.2 ± 0.7, and -0.1 ± 0.3 mm in vertical, longitudinal, and lateral direction. For the thoracic case, the mean differences were 0.3 ± 0.5, -0.6 ± 1.9, 0.0 ± 0.4 mm. CONCLUSIONS: The presented system copes with the increasing demand for more accurate patient positioning due to more precise irradiation technologies and minimizes the preparation times for correct patient alignment, therefore optimizing the treatment workflow. Moreover, TOPOS is a versatile and cost effective image guided radiation therapy device. It allows an objective rating of the patient's position before and during the irradiation and could also be used for respiratory gating or tracking.

PURPOSE: A patient positioning system for radiation therapy based on structured white light and using off-the-shelf hardware components for flexibility and cost-effectiveness has been developed in house. Increased accuracy, patient comfort, abandonment of any skin marks, accelerated workflow, objective reading/recording, better usability and robust sensor design, compared to other positioning approaches, were the main goals of this work. Another aim was the application of a 6 degrees of freedom tracking system working without dose deposition. METHODS: Two optical sensors are the main parts of the TOPOS® system (Topometrical Positioning, cyberTECHNOLOGIES, Germany). The components: cameras, projectors, and computers are commercial off-the-shelf products, allowing for low production costs. The black/white cameras of the prototype are capable of taking up to 240 frames per second (resolution: 640 × 488 pixels). The projector has a resolution of 1024 × 768 and a refresh rate of 120 Hz. The patient's body surface is measured continuously and registered to a reference surface, providing a transformation to superimpose the patient's surface to the reference (planning CT) surface as best as possible. The execution of the calculated transformation provides the correct patient position before the treatment starts. Due to the high-speed acquisition of the surfaces, a surveillance of the patient's (respiration) motion during treatment is also accomplished. The accuracy of the system was determined using a male mannequin. Two treatment sites were evaluated: one simulating a head and neck treatment and the other simulating a thoracic wall treatment. The mannequin was moved to predefined positions, and shift vectors given by the surface registration were evaluated. Additionally manual positioning using a color-coding system was evaluated. RESULTS: Two prototypes have been developed, each allowing a continuous high density scan of a 500 × 500 × 400 mm(3) (L × W × D) large volume with a refresh rate of 10 Hz (extendible to 20 Hz for a single sensor system). Surface and position correction display, as well as respiratory motion, is shown in real-time (delay < 200 ms) using present graphical hardware acceleration. For an intuitive view of the patient's misalignment, a fast surface registration algorithm has been developed and tested and a real-time color-coding technique is proposed and verified that allows the user to easily verify the position of the patient. Using first the surface registration and then the color coding the best results were obtained: for the head and neck case, the mean difference between the actual zero position and the final match was 0.1 ± 0.4, -0.2 ± 0.7, and -0.1 ± 0.3 mm in vertical, longitudinal, and lateral direction. For the thoracic case, the mean differences were 0.3 ± 0.5, -0.6 ± 1.9, 0.0 ± 0.4 mm. CONCLUSIONS: The presented system copes with the increasing demand for more accurate patient positioning due to more precise irradiation technologies and minimizes the preparation times for correct patient alignment, therefore optimizing the treatment workflow. Moreover, TOPOS is a versatile and cost effective image guided radiation therapy device. It allows an objective rating of the patient's position before and during the irradiation and could also be used for respiratory gating or tracking.

Citations

4 citations in Web of Science®
5 citations in Scopus®
Google Scholar™

Altmetrics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Radiation Oncology
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:2013
Deposited On:19 Jun 2013 13:52
Last Modified:05 Apr 2016 16:49
Publisher:American Association of Physicists in Medicine
ISSN:0094-2405
Publisher DOI:https://doi.org/10.1118/1.4794927
PubMed ID:23556922

Download

Full text not available from this repository.View at publisher

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.

Author Collaborations