Evaluation of a continuous glucose monitoring system compared with an in-house standard laboratory assay and a handheld point-of-care glucometer in critically ill neonatal foals


Hug, Sophie A; Riond, Barbara; Schwarzwald, Colin C (2013). Evaluation of a continuous glucose monitoring system compared with an in-house standard laboratory assay and a handheld point-of-care glucometer in critically ill neonatal foals. Journal of Veterinary Emergency and Critical Care, 23(4):408-415.

Abstract

OBJECTIVE: To evaluate the applicability and accuracy of a continuous glucose monitoring system (CGMS) in critically ill foals by comparing the performance of the CGMS, a point-of-care (POC) glucometer, and an in-house standard laboratory assay (SLA).
DESIGN: Prospective study.
SETTING: University teaching hospital.
ANIMALS: Seven critically ill neonatal foals requiring intensive care.
INTERVENTIONS: Foals were instrumented with a CGMS that measured interstitial glucose concentration every 5 minutes for the duration of 15-79 hours. Capillary and venous blood samples were taken every 4-6 hours for POC and SLA measurements, respectively.
MEASUREMENTS AND MAIN RESULTS: Bland-Altman analysis showed a mean bias (95% limits of agreement) of -0.1 (-3.9 to 3.5) mmol/L for comparison of CGMS versus SLA, 0.06 (-3.9 to 4.0) mmol/L for comparison of CGMS versus POC glucometer, and -0.16 (-1.8 to 1.5) mmol/L for comparison of POC glucometer versus SLA. Percent agreement and weighted kappa for classification in hypoglycemia, normoglycemia, and hyperglycemia were 68.4% and 0.296 for CGMS versus SLA, 72.4% and 0.442 for CGMS versus POC glucometer, and 80.7% and 0.568 for POC glucometer versus SLA.
CONCLUSIONS: The CGMS may be helpful for monitoring a trend in interstitial glucose concentration in critically ill neonatal foals. However, considering the wide limits of agreement between methods, the CGMS should only be used as an adjunctive device to other, more accurate and readily available methods that are able to detect acute changes in glucose concentration. Its use is further limited by the relatively high costs of the sensors, the mandatory 2-hour initialization period, and the difficulties of keeping the transducer in place in an active foal. The POC glucometer used in this study is easy to use and proved to be sufficiently accurate for repeated, stall-sided glucose monitoring in neonatal foals.

OBJECTIVE: To evaluate the applicability and accuracy of a continuous glucose monitoring system (CGMS) in critically ill foals by comparing the performance of the CGMS, a point-of-care (POC) glucometer, and an in-house standard laboratory assay (SLA).
DESIGN: Prospective study.
SETTING: University teaching hospital.
ANIMALS: Seven critically ill neonatal foals requiring intensive care.
INTERVENTIONS: Foals were instrumented with a CGMS that measured interstitial glucose concentration every 5 minutes for the duration of 15-79 hours. Capillary and venous blood samples were taken every 4-6 hours for POC and SLA measurements, respectively.
MEASUREMENTS AND MAIN RESULTS: Bland-Altman analysis showed a mean bias (95% limits of agreement) of -0.1 (-3.9 to 3.5) mmol/L for comparison of CGMS versus SLA, 0.06 (-3.9 to 4.0) mmol/L for comparison of CGMS versus POC glucometer, and -0.16 (-1.8 to 1.5) mmol/L for comparison of POC glucometer versus SLA. Percent agreement and weighted kappa for classification in hypoglycemia, normoglycemia, and hyperglycemia were 68.4% and 0.296 for CGMS versus SLA, 72.4% and 0.442 for CGMS versus POC glucometer, and 80.7% and 0.568 for POC glucometer versus SLA.
CONCLUSIONS: The CGMS may be helpful for monitoring a trend in interstitial glucose concentration in critically ill neonatal foals. However, considering the wide limits of agreement between methods, the CGMS should only be used as an adjunctive device to other, more accurate and readily available methods that are able to detect acute changes in glucose concentration. Its use is further limited by the relatively high costs of the sensors, the mandatory 2-hour initialization period, and the difficulties of keeping the transducer in place in an active foal. The POC glucometer used in this study is easy to use and proved to be sufficiently accurate for repeated, stall-sided glucose monitoring in neonatal foals.

Citations

Altmetrics

Downloads

1 download since deposited on 26 Jul 2013
0 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:05 Vetsuisse Faculty > Veterinary Clinic > Equine Department
05 Vetsuisse Faculty > Veterinary Clinic > Department of Farm Animals
Dewey Decimal Classification:570 Life sciences; biology
630 Agriculture
Language:English
Date:2013
Deposited On:26 Jul 2013 12:27
Last Modified:05 Apr 2016 16:53
Publisher:Wiley-Blackwell
ISSN:1476-4431
Publisher DOI:https://doi.org/10.1111/vec.12072
PubMed ID:23859299
Permanent URL: https://doi.org/10.5167/uzh-79545

Download

[img]
Content: Published Version
Filetype: PDF - Registered users only
Size: 418kB
View at publisher

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.

Author Collaborations