Automated urinalysis: first experiences and a comparison between the Iris iQ200 urine microscopy system, the Sysmex UF-100 flow cytometer and manual microscopic particle counting


Shayanfar, Noushin; Tobler, Ulrich; von Eckardstein, Arnold; Bestmann, Lukas (2007). Automated urinalysis: first experiences and a comparison between the Iris iQ200 urine microscopy system, the Sysmex UF-100 flow cytometer and manual microscopic particle counting. Clinical Chemistry and Laboratory Medicine, 45(9):1251-1256.

Abstract

BACKGROUND: Automated analysis of insoluble urine components can reduce the workload of conventional microscopic examination of urine sediment and is possibly helpful for standardization. We compared the diagnostic performance of two automated urine sediment analyzers and combined dipstick/automated urine analysis with that of the traditional dipstick/microscopy algorithm.
METHODS: A total of 332 specimens were collected and analyzed for insoluble urine components by microscopy and automated analyzers, namely the Iris iQ200 (Iris Diagnostics) and the UF-100 flow cytometer (Sysmex).
RESULTS: The coefficients of variation for day-to-day quality control of the iQ200 and UF-100 analyzers were 6.5% and 5.5%, respectively, for red blood cells. We reached accuracy ranging from 68% (bacteria) to 97% (yeast) for the iQ200 and from 42% (bacteria) to 93% (yeast) for the UF-100. The combination of dipstick and automated urine sediment analysis increased the sensitivity of screening to approximately 98%.
CONCLUSIONS: We conclude that automated urine sediment analysis is sufficiently precise and improves the workflow in a routine laboratory. In addition, it allows sediment analysis of all urine samples and thereby helps to detect pathological samples that would have been missed in the conventional two-step procedure according to the European guidelines. Although it is not a substitute for microscopic sediment examination, it can, when combined with dipstick testing, reduce the number of specimens submitted to microscopy. Visual microscopy is still required for some samples, namely, dysmorphic erythrocytes, yeasts, Trichomonas, oval fat bodies, differentiation of casts and certain crystals.

BACKGROUND: Automated analysis of insoluble urine components can reduce the workload of conventional microscopic examination of urine sediment and is possibly helpful for standardization. We compared the diagnostic performance of two automated urine sediment analyzers and combined dipstick/automated urine analysis with that of the traditional dipstick/microscopy algorithm.
METHODS: A total of 332 specimens were collected and analyzed for insoluble urine components by microscopy and automated analyzers, namely the Iris iQ200 (Iris Diagnostics) and the UF-100 flow cytometer (Sysmex).
RESULTS: The coefficients of variation for day-to-day quality control of the iQ200 and UF-100 analyzers were 6.5% and 5.5%, respectively, for red blood cells. We reached accuracy ranging from 68% (bacteria) to 97% (yeast) for the iQ200 and from 42% (bacteria) to 93% (yeast) for the UF-100. The combination of dipstick and automated urine sediment analysis increased the sensitivity of screening to approximately 98%.
CONCLUSIONS: We conclude that automated urine sediment analysis is sufficiently precise and improves the workflow in a routine laboratory. In addition, it allows sediment analysis of all urine samples and thereby helps to detect pathological samples that would have been missed in the conventional two-step procedure according to the European guidelines. Although it is not a substitute for microscopic sediment examination, it can, when combined with dipstick testing, reduce the number of specimens submitted to microscopy. Visual microscopy is still required for some samples, namely, dysmorphic erythrocytes, yeasts, Trichomonas, oval fat bodies, differentiation of casts and certain crystals.

Citations

24 citations in Web of Science®
33 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

1098 downloads since deposited on 19 Nov 2013
410 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Institute of Clinical Chemistry
Dewey Decimal Classification:610 Medicine & health
540 Chemistry
Language:English
Date:2007
Deposited On:19 Nov 2013 15:53
Last Modified:05 Apr 2016 17:02
Publisher:De Gruyter
ISSN:1434-6621
Publisher DOI:https://doi.org/10.1515/CCLM.2007.503
PubMed ID:17635081
Permanent URL: https://doi.org/10.5167/uzh-81721

Download

[img]
Preview
Content: Published Version
Filetype: PDF
Size: 98kB
View at publisher

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.

Author Collaborations