Lateral temperature spread of monopolar, bipolar and ultrasonic instruments for robotic assisted laparoscopic surgery


Hefermehl, Lukas J; Largo, Remo A; Hermanns, Thomas; Poyet, Cédric; Sulser, Tullio; Eberli, Daniel (2014). Lateral temperature spread of monopolar, bipolar and ultrasonic instruments for robotic assisted laparoscopic surgery. BJU International, 114(2):245-252.

Abstract

OBJECTIVE: To assess cautery instruments used in robotic assisted laparoscopic (RAL) surgery in regard to critical heat spread on tissue and cellular level.
MATERIALS AND METHODS: Thermal spread along bovine musculofascial tissues was examined by infrared camera, histology and enzyme assay. Currently used monopolar, bipolar and ultrasonic laparoscopic instruments were investigated at various power settings and application times. The efficacy of using an additional Maryland clamp as a heat sink was evaluated. 45°C was considered the cutoff temperature for possible nerve damage.
RESULTS: Monopolar instruments exhibited a critical spread of 3.5 ± 2.3 mm when applied at 60 W for 1 s. After 2 s, the spread surpassed 20 mm. For adjustable bipolar instruments the spread was 2.2 ± 0.6 mm at 60 W and 1 s and 3.6 ± 1.3 mm at 2 s. PK and LigaSure forceps had spreads of 3.9 ± 0.8 mm and 2.8 ± 0.6 mm respectively, whereas the ultrasonic instrument reached 2.9 ± 0.8 mm. Application of an additional Maryland clamp as a heat sink significantly reduced the thermal spread. Histomorphometric analyses and enzyme assay supported these findings.
CONCLUSION: All coagulation devices used in robotic assisted laparoscopy surgery have distinct thermal spreads depending on power setting and application time Cautery may be of concern due to lateral temperature spread, causing potential damage to sensitive structures including nerves. Our results provide surgeons with a resource for educated decision making when using coagulation devices during robotic procedures.

Abstract

OBJECTIVE: To assess cautery instruments used in robotic assisted laparoscopic (RAL) surgery in regard to critical heat spread on tissue and cellular level.
MATERIALS AND METHODS: Thermal spread along bovine musculofascial tissues was examined by infrared camera, histology and enzyme assay. Currently used monopolar, bipolar and ultrasonic laparoscopic instruments were investigated at various power settings and application times. The efficacy of using an additional Maryland clamp as a heat sink was evaluated. 45°C was considered the cutoff temperature for possible nerve damage.
RESULTS: Monopolar instruments exhibited a critical spread of 3.5 ± 2.3 mm when applied at 60 W for 1 s. After 2 s, the spread surpassed 20 mm. For adjustable bipolar instruments the spread was 2.2 ± 0.6 mm at 60 W and 1 s and 3.6 ± 1.3 mm at 2 s. PK and LigaSure forceps had spreads of 3.9 ± 0.8 mm and 2.8 ± 0.6 mm respectively, whereas the ultrasonic instrument reached 2.9 ± 0.8 mm. Application of an additional Maryland clamp as a heat sink significantly reduced the thermal spread. Histomorphometric analyses and enzyme assay supported these findings.
CONCLUSION: All coagulation devices used in robotic assisted laparoscopy surgery have distinct thermal spreads depending on power setting and application time Cautery may be of concern due to lateral temperature spread, causing potential damage to sensitive structures including nerves. Our results provide surgeons with a resource for educated decision making when using coagulation devices during robotic procedures.

Citations

8 citations in Web of Science®
7 citations in Scopus®
Google Scholar™

Altmetrics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Urological Clinic
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:2014
Deposited On:28 Oct 2013 16:20
Last Modified:05 Apr 2016 17:04
Publisher:Wiley-Blackwell
ISSN:1464-4096
Free access at:Publisher DOI. An embargo period may apply.
Publisher DOI:https://doi.org/10.1111/bju.12498
PubMed ID:24127773

Download

Full text not available from this repository.
View at publisher

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.

Author Collaborations