Multicenter audiometric results with the Vibrant Soundbridge, a semi-implantable hearing device for sensorineural hearing impairment


Snik, A F; Mylanus, E A; Cremers, C W; Dillier, N; Fisch, U; Gnadeberg, D; Lenarz, T; Mazolli, M; Babighian, G; Uziel, A S; Cooper, H R; O'Connor, A F; Fraysse, B; Charachon, R; Shehata-Dieler, W E (2001). Multicenter audiometric results with the Vibrant Soundbridge, a semi-implantable hearing device for sensorineural hearing impairment. Otolaryngologic Clinics of North America, 34(2):373-388.

Abstract

The Vibrant Soundbridge, a semi-implantable hearing device for subjects with moderate to severe sensorineural hearing impairment was introduced commercially. First audiologic results are presented on 63 patients from 10 European implant centers. Hearing loss was at 0.5, 1, 2, and 4 kHz varying between 43 and 81 dB HL. The patients used the analogue audio processor, type 302. Measured sound-field gain was compared with NAL-R target values. For most patients an acceptable agreement was found. There was a subgroup of patients, however, with relatively low gain. The results suggest that this was related to the suboptimal positioning and fixation of the transducer to the incus.

Abstract

The Vibrant Soundbridge, a semi-implantable hearing device for subjects with moderate to severe sensorineural hearing impairment was introduced commercially. First audiologic results are presented on 63 patients from 10 European implant centers. Hearing loss was at 0.5, 1, 2, and 4 kHz varying between 43 and 81 dB HL. The patients used the analogue audio processor, type 302. Measured sound-field gain was compared with NAL-R target values. For most patients an acceptable agreement was found. There was a subgroup of patients, however, with relatively low gain. The results suggest that this was related to the suboptimal positioning and fixation of the transducer to the incus.

Citations

43 citations in Web of Science®
56 citations in Scopus®
Google Scholar™

Altmetrics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Otorhinolaryngology
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:2001
Deposited On:30 Mar 2009 13:08
Last Modified:05 Apr 2016 12:43
Publisher:Harcourt Brace Jovanovich / W B Saunders
ISSN:0030-6665
PubMed ID:11382576

Download

Full text not available from this repository.

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.

Author Collaborations