UZH-Logo

Maintenance Infos

Intraoperative assessment of stapes movement


Huber, A; Linder, T; Ferrazzini, M; Schmid, S; Dillier, N; Stoeckli, S; Fisch, U (2001). Intraoperative assessment of stapes movement. Annals of Otology, Rhinology and Laryngology, 110(1):31-35.

Abstract

A method is described that allows, for the first time, intraoperative vibration modes assessment of the acoustically stimulated stapes by means of scanning laser Doppler interferometry (LDI). The study was designed to answer the following questions: 1) Is LDI practical for taking measurements during surgery? 2) Are the results comparable to the findings in temporal bone preparations? and 3) Do the vibration characteristics of the stapes change after the posterior incudal ligament is detached from the incus? Seven patients with profound bilateral hearing loss who were undergoing cochlear implantation were included in the study. The measurement system was easily applicable for intraoperative measurements and allowed contact-free analysis with very high accuracy. No major differences in the results from the live human subjects and temporal bone preparations were observed. The stapes movement was predominantly pistonlike at the lower frequencies and became complex at higher frequencies. Sacrificing the posterior incudal ligament had no statistically significant effect on stapes vibration.

Abstract

A method is described that allows, for the first time, intraoperative vibration modes assessment of the acoustically stimulated stapes by means of scanning laser Doppler interferometry (LDI). The study was designed to answer the following questions: 1) Is LDI practical for taking measurements during surgery? 2) Are the results comparable to the findings in temporal bone preparations? and 3) Do the vibration characteristics of the stapes change after the posterior incudal ligament is detached from the incus? Seven patients with profound bilateral hearing loss who were undergoing cochlear implantation were included in the study. The measurement system was easily applicable for intraoperative measurements and allowed contact-free analysis with very high accuracy. No major differences in the results from the live human subjects and temporal bone preparations were observed. The stapes movement was predominantly pistonlike at the lower frequencies and became complex at higher frequencies. Sacrificing the posterior incudal ligament had no statistically significant effect on stapes vibration.

Citations

65 citations in Web of Science®
68 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

34 downloads since deposited on 30 Mar 2009
13 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Otorhinolaryngology
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:2001
Deposited On:30 Mar 2009 11:44
Last Modified:05 Apr 2016 12:43
Publisher:Annals Publishing
ISSN:0003-4894
Publisher DOI:https://doi.org/10.1177/000348940111000106
PubMed ID:11201805

Download

[img]
Preview
Filetype: PDF
Size: 249kB
View at publisher

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.