Analysis of the best site on the stapes footplate for ossicular chain reconstruction


Asai, M; Huber, A M; Goode, R L (1999). Analysis of the best site on the stapes footplate for ossicular chain reconstruction. Acta Oto-Laryngologica, 119(3):356-361.

Abstract

Experiments were performed in 22 fresh human temporal bones to compare the relative acoustic function of three stapes footplate sites for an incus stapes superstructure replacement prosthesis (I-SRP). The three sites evaluated were the anterior, centre and posterior footplates. A new round window (RW) measurement method was used to make the comparisons. A small glass microsphere was placed in the centre of the RW as a target. A Polytec laser Doppler vibrometer was used to measure round window displacement in response to 50 pure tones between 200 and 10,000 Hz presented at 80 dB SPL at the tympanic membrane (TM). After a baseline measurement of RW displacement in the intact temporal bone, the incus was removed and a cement I-SRP (CIRP) formed between the mid-malleus handle and each of the three test footplate sites, in random order. RW displacement was again measured after placement of the CIRP at each of three sites. We found the centre site to be 3.0-7.0 dB better than the anterior site above 2,000 Hz. There were no differences between the anterior and centre sites below 2,000 Hz. The posterior site was the worst at all frequencies.

Experiments were performed in 22 fresh human temporal bones to compare the relative acoustic function of three stapes footplate sites for an incus stapes superstructure replacement prosthesis (I-SRP). The three sites evaluated were the anterior, centre and posterior footplates. A new round window (RW) measurement method was used to make the comparisons. A small glass microsphere was placed in the centre of the RW as a target. A Polytec laser Doppler vibrometer was used to measure round window displacement in response to 50 pure tones between 200 and 10,000 Hz presented at 80 dB SPL at the tympanic membrane (TM). After a baseline measurement of RW displacement in the intact temporal bone, the incus was removed and a cement I-SRP (CIRP) formed between the mid-malleus handle and each of the three test footplate sites, in random order. RW displacement was again measured after placement of the CIRP at each of three sites. We found the centre site to be 3.0-7.0 dB better than the anterior site above 2,000 Hz. There were no differences between the anterior and centre sites below 2,000 Hz. The posterior site was the worst at all frequencies.

Citations

52 citations in Web of Science®
54 citations in Scopus®
Google Scholar™

Altmetrics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Otorhinolaryngology
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:1999
Deposited On:30 Mar 2009 10:33
Last Modified:05 Apr 2016 12:43
Publisher:Taylor & Francis
ISSN:0001-6489
PubMed ID:10380743

Download

Full text not available from this repository.

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.

Author Collaborations