Primary stability and stiffness in ankle arthrodes-crossed screws versus anterior plating


Betz, Michael M; Benninger, Emanuel E; Favre, Philipp P; Wieser, Karl K; Vich, Magdalena M; Espinosa, Norman (2013). Primary stability and stiffness in ankle arthrodes-crossed screws versus anterior plating. Foot and Ankle Surgery, 19(3):168-172.

Abstract

BACKGROUND: Ankle arthrodesis is commonly used for the treatment of osteoarthritis or failed arthroplasty. Screw fixation is the predominant technique to perform ankle arthrodesis. Due to a considerable frequency of failures research suggests the use of an anatomically shaped anterior double plate system as a reliable method for isolated tibiotalar arthrodesis. The purpose of the present biomechanical study was to compare two groups of ankle fusion constructs - three screw fixation and an anterior double plate system - in terms of primary stability and stiffness.
METHODS: Six matched-pairs human cadaveric lower legs (Thiel fixated) were used in this study. One specimen from each pair was randomly assigned to be stabilized with the anterior double plate system and the other with the three-screw technique. The different arthrodesis methods were tested by dorsiflexing the foot until failure of the system, defined as rotation of the talus relative to the tibia in the sagittal plane. Experiments were performed on a universal materials testing machine. The force required to make arthrodesis fail was documented. For calculation of the stiffness, a linear regression was fitted to the force-displacement curve in the linear portion of the curve and its slope taken as the stiffness.
RESULTS: For the anatomically shaped double-plate system a mean load of 967N was needed (range from 570N to 1400N) to make arthrodesis fail. The three-screw fixation method resisted a mean load of 190N (range from 100N to 280N) (p=0.005). In terms of stiffness a mean of 56N/mm (range from 35N/mm to 79N/mm) was achieved for the anatomically shaped double-plate system whereas a mean of 10N/mm (range from 6N/mm to 18N/mm) was achieved for the three-screw fixation method (p=0.004).
CONCLUSIONS: Our biomechanical data demonstrates that the anterior double-plate system is significantly superior to the three-screw fixation technique for ankle arthrodesis in terms of primary stability and stiffness.

Abstract

BACKGROUND: Ankle arthrodesis is commonly used for the treatment of osteoarthritis or failed arthroplasty. Screw fixation is the predominant technique to perform ankle arthrodesis. Due to a considerable frequency of failures research suggests the use of an anatomically shaped anterior double plate system as a reliable method for isolated tibiotalar arthrodesis. The purpose of the present biomechanical study was to compare two groups of ankle fusion constructs - three screw fixation and an anterior double plate system - in terms of primary stability and stiffness.
METHODS: Six matched-pairs human cadaveric lower legs (Thiel fixated) were used in this study. One specimen from each pair was randomly assigned to be stabilized with the anterior double plate system and the other with the three-screw technique. The different arthrodesis methods were tested by dorsiflexing the foot until failure of the system, defined as rotation of the talus relative to the tibia in the sagittal plane. Experiments were performed on a universal materials testing machine. The force required to make arthrodesis fail was documented. For calculation of the stiffness, a linear regression was fitted to the force-displacement curve in the linear portion of the curve and its slope taken as the stiffness.
RESULTS: For the anatomically shaped double-plate system a mean load of 967N was needed (range from 570N to 1400N) to make arthrodesis fail. The three-screw fixation method resisted a mean load of 190N (range from 100N to 280N) (p=0.005). In terms of stiffness a mean of 56N/mm (range from 35N/mm to 79N/mm) was achieved for the anatomically shaped double-plate system whereas a mean of 10N/mm (range from 6N/mm to 18N/mm) was achieved for the three-screw fixation method (p=0.004).
CONCLUSIONS: Our biomechanical data demonstrates that the anterior double-plate system is significantly superior to the three-screw fixation technique for ankle arthrodesis in terms of primary stability and stiffness.

Citations

4 citations in Web of Science®
4 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

72 downloads since deposited on 09 Dec 2013
29 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > Balgrist University Hospital, Swiss Spinal Cord Injury Center
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:September 2013
Deposited On:09 Dec 2013 08:01
Last Modified:05 Apr 2016 17:13
Publisher:Elsevier
ISSN:1268-7731
Publisher DOI:https://doi.org/10.1016/j.fas.2013.04.006
PubMed ID:23830164

Download

[img]
Preview
Content: Accepted Version
Filetype: PDF
Size: 149kB
View at publisher

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.

Author Collaborations