Influence of residual bone thickness on primary stability of hybrid self-tapping and cylindric non-self-tapping implants in vitro


Divac, Marija; Stawarczyk, Bogna; Sahrmann, Philipp; Attin, Thomas; Schmidlin, Patrick R (2013). Influence of residual bone thickness on primary stability of hybrid self-tapping and cylindric non-self-tapping implants in vitro. International Journal of Oral & Maxillofacial Implants, 28(1):84-88.

Abstract

PURPOSE: To assess the primary stability of a hybrid self-tapping implant and a cylindric non-self-tapping implant in an in vitro test model using polyurethane foam. MATERIALS AND METHODS: Eighty standardized blocks of cellular rigid polyurethane foam, 2 cm long and 1 cm wide, with different thicknesses of 2, 4, 6, and 9 mm (n = 10 per group) were cut. Two implant systems--a hybrid self-tapping (Tapered Effect [TE], Straumann) and a cylindric non-self-tapping (Standard Plus [SP] Wide Neck, Straumann) were placed in the block specimens. Subsequently, resonance frequency analysis (RFA) was performed. The RFA measurements were made in triplicate on four aspects of each implant (mesial, distal, buccal, and oral), and the mean RFA value was calculated. Subsequently, the tensile load of the implants was determined by pull-out tests. The data were analyzed using one-way and two-way analysis of variance followed by a post hoc Scheffe test and a t test (α = .05). Additionally, the simple linear correlation between the RFA and tensile load values was evaluated. RESULTS: No statistically significant differences were found between TE and SP in terms of RFA at different bone thicknesses. Starting from a bone thickness of 4 mm, TE implants showed significantly higher tensile load compared to SP implants (P = .016 to .040). A correlation was found between the RFA measurements and tensile load. CONCLUSIONS: Mechanically stable placement is possible with TE and SP implants in a trabecular bone model. RFA and tensile load increased with greater bone thickness.

PURPOSE: To assess the primary stability of a hybrid self-tapping implant and a cylindric non-self-tapping implant in an in vitro test model using polyurethane foam. MATERIALS AND METHODS: Eighty standardized blocks of cellular rigid polyurethane foam, 2 cm long and 1 cm wide, with different thicknesses of 2, 4, 6, and 9 mm (n = 10 per group) were cut. Two implant systems--a hybrid self-tapping (Tapered Effect [TE], Straumann) and a cylindric non-self-tapping (Standard Plus [SP] Wide Neck, Straumann) were placed in the block specimens. Subsequently, resonance frequency analysis (RFA) was performed. The RFA measurements were made in triplicate on four aspects of each implant (mesial, distal, buccal, and oral), and the mean RFA value was calculated. Subsequently, the tensile load of the implants was determined by pull-out tests. The data were analyzed using one-way and two-way analysis of variance followed by a post hoc Scheffe test and a t test (α = .05). Additionally, the simple linear correlation between the RFA and tensile load values was evaluated. RESULTS: No statistically significant differences were found between TE and SP in terms of RFA at different bone thicknesses. Starting from a bone thickness of 4 mm, TE implants showed significantly higher tensile load compared to SP implants (P = .016 to .040). A correlation was found between the RFA measurements and tensile load. CONCLUSIONS: Mechanically stable placement is possible with TE and SP implants in a trabecular bone model. RFA and tensile load increased with greater bone thickness.

Citations

2 citations in Web of Science®
2 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

2 downloads since deposited on 13 Dec 2013
0 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > Center for Dental Medicine > Clinic for Preventive Dentistry, Periodontology and Cariology
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:2013
Deposited On:13 Dec 2013 14:25
Last Modified:05 Apr 2016 17:14
Publisher:Quintessence Publishing
ISSN:0882-2786
Publisher DOI:https://doi.org/10.11607/jomi.2587
Related URLs:http://www.zora.uzh.ch/79931/
PubMed ID:23377051
Permanent URL: https://doi.org/10.5167/uzh-86323

Download

[img]
Content: Accepted Version
Language: English
Filetype: PDF - Registered users only
Size: 809kB
View at publisher

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.

Author Collaborations