Fracture load of milled polymeric fixed dental prostheses as a function of connector cross-sectional areas


Wimmer, Timea; Ender, Andreas; Roos, Malgorzata; Stawarczyk, Bogna (2013). Fracture load of milled polymeric fixed dental prostheses as a function of connector cross-sectional areas. Journal of Prosthetic Dentistry, 110(4):288-295.

Abstract

STATEMENT OF PROBLEM: Polymeric computer-aided design/computer-aided manufacturing (CAD/CAM) blocks exhibit various advantages in contrast to conventionally processed resin restorations. However, the influence of connector dimensions on resin fixed dental prostheses (FDPs) has not yet been investigated.
PURPOSE: This study evaluated the impact of connector cross-sectional area (CSA) on the fracture load of 3-unit CAD/CAM FDPs and compared this with conventionally fabricated ones.
MATERIAL AND METHODS: Anatomically shaped 3-unit FDPs with the CSAs of 6, 9, 12, and 16 mm(2) (N=240, n=15 per material and per CSA) were fabricated from the following CAD/CAM materials: artBloc Temp (AT), TelioCAD (TC), CAD-Temp (CT), and one conventional resin material, CronMix K (CMK) as the control group. The fracture load was tested and the data were analyzed with 2-way ANOVA, 1-way ANOVA, the Scheffé post hoc test, and Weibull statistics (α=05).
RESULTS: CMK showed the significant lowest values for all CSAs followed by CT, except for the 12 mm(2) connector (P<.001). The CAD/CAM FDPs exhibited a significant increase in fracture load with the increase of CSA (P<.001). Conventionally fabricated FDPs presented an increase of values up to the CSA of 12 mm(2). For TC, the shape increased with a larger CSA. The other materials exhibited no tendencies in this regard. Among the 12 mm(2) groups, AT exhibited the highest shape value (19.1), and among the 16 mm(2) groups, TC showed the highest shape value (17.0). The CMK FDPs with a CSA of 16 mm(2) showed almost half as high a shape parameter (6.4) than the other three CSAs.
CONCLUSIONS: CAD/CAM resin FDPs revealed significantly higher fracture load values than conventionally fabricated FDPs and showed a significant increase in fracture load with the increase of the CSA.

STATEMENT OF PROBLEM: Polymeric computer-aided design/computer-aided manufacturing (CAD/CAM) blocks exhibit various advantages in contrast to conventionally processed resin restorations. However, the influence of connector dimensions on resin fixed dental prostheses (FDPs) has not yet been investigated.
PURPOSE: This study evaluated the impact of connector cross-sectional area (CSA) on the fracture load of 3-unit CAD/CAM FDPs and compared this with conventionally fabricated ones.
MATERIAL AND METHODS: Anatomically shaped 3-unit FDPs with the CSAs of 6, 9, 12, and 16 mm(2) (N=240, n=15 per material and per CSA) were fabricated from the following CAD/CAM materials: artBloc Temp (AT), TelioCAD (TC), CAD-Temp (CT), and one conventional resin material, CronMix K (CMK) as the control group. The fracture load was tested and the data were analyzed with 2-way ANOVA, 1-way ANOVA, the Scheffé post hoc test, and Weibull statistics (α=05).
RESULTS: CMK showed the significant lowest values for all CSAs followed by CT, except for the 12 mm(2) connector (P<.001). The CAD/CAM FDPs exhibited a significant increase in fracture load with the increase of CSA (P<.001). Conventionally fabricated FDPs presented an increase of values up to the CSA of 12 mm(2). For TC, the shape increased with a larger CSA. The other materials exhibited no tendencies in this regard. Among the 12 mm(2) groups, AT exhibited the highest shape value (19.1), and among the 16 mm(2) groups, TC showed the highest shape value (17.0). The CMK FDPs with a CSA of 16 mm(2) showed almost half as high a shape parameter (6.4) than the other three CSAs.
CONCLUSIONS: CAD/CAM resin FDPs revealed significantly higher fracture load values than conventionally fabricated FDPs and showed a significant increase in fracture load with the increase of the CSA.

Citations

3 citations in Web of Science®
3 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

3 downloads since deposited on 14 Jan 2014
0 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > Epidemiology, Biostatistics and Prevention Institute (EBPI)
04 Faculty of Medicine > Center for Dental Medicine > Clinic for Preventive Dentistry, Periodontology and Cariology
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:2013
Deposited On:14 Jan 2014 15:46
Last Modified:05 Apr 2016 17:23
Publisher:Mosby, Inc.
ISSN:0022-3913
Publisher DOI:https://doi.org/10.1016/S0022-3913(13)60378-9
PubMed ID:24079564
Permanent URL: https://doi.org/10.5167/uzh-88318

Download

[img]
Content: Published Version
Language: English
Filetype: PDF - Registered users only
Size: 201kB
View at publisher
[img]
Content: Accepted Version
Filetype: PDF - Registered users only
Size: 603kB

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.

Author Collaborations