Safety and feasibility of novel technology fusing echocardiography and fluoroscopy images during MitraClip interventions


Sündermann, Simon H; Biaggi, Patric; Grünenfelder, Jürg; Gessat, Michael; Felix, Christian; Bettex, Dominique; Falk, Volkmar; Corti, Roberto (2014). Safety and feasibility of novel technology fusing echocardiography and fluoroscopy images during MitraClip interventions. EuroIntervention, 9(10):1210-1216.

Abstract

Aims: The EchoNavigator (EN) software (Philips Healthcare, Best, The Netherlands) enables real-time fusion of echocardiography and fluoroscopy by co-registration of the echocardiography probe on the x-ray image. We aimed to evaluate the feasibility and safety of this novel software during MitraClip procedures. Methods and results: Twenty-one patients were treated with the support of EchoNavigator software (EN+ patients). The primary (safety) endpoint was the total radiation dose. Secondary endpoints were fluoroscopy and total procedure time. The measurements were compared to those of 21 patients treated immediately before the installation of EchoNavigator (EN- patients). More MitraClips (45 vs. 36) were implanted in the EN+ group, mirroring more complex interventions in this group. In EN+ patients, radiation dose (Gy/cm(2)) was similar compared to EN- patients (146.5±123.6 vs.146.8±134.1, p=0.9). Total procedure time (minutes) was similar in the EN+ group compared to EN- patients (136.2±50.2 vs. 125.7±51.2, p=0.5). The main benefit of the EchoNavigator is the automated real-time fusion of echocardiography and fluoroscopy, leading to easier catheter manipulation. Conclusions: The use of EchoNavigator software was feasible and safe in all study patients. Further studies are necessary to confirm the benefits of using this software.

Abstract

Aims: The EchoNavigator (EN) software (Philips Healthcare, Best, The Netherlands) enables real-time fusion of echocardiography and fluoroscopy by co-registration of the echocardiography probe on the x-ray image. We aimed to evaluate the feasibility and safety of this novel software during MitraClip procedures. Methods and results: Twenty-one patients were treated with the support of EchoNavigator software (EN+ patients). The primary (safety) endpoint was the total radiation dose. Secondary endpoints were fluoroscopy and total procedure time. The measurements were compared to those of 21 patients treated immediately before the installation of EchoNavigator (EN- patients). More MitraClips (45 vs. 36) were implanted in the EN+ group, mirroring more complex interventions in this group. In EN+ patients, radiation dose (Gy/cm(2)) was similar compared to EN- patients (146.5±123.6 vs.146.8±134.1, p=0.9). Total procedure time (minutes) was similar in the EN+ group compared to EN- patients (136.2±50.2 vs. 125.7±51.2, p=0.5). The main benefit of the EchoNavigator is the automated real-time fusion of echocardiography and fluoroscopy, leading to easier catheter manipulation. Conclusions: The use of EchoNavigator software was feasible and safe in all study patients. Further studies are necessary to confirm the benefits of using this software.

Citations

7 citations in Web of Science®
11 citations in Scopus®
Google Scholar™

Altmetrics

Additional indexing

Item Type:Journal Article, not refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Cardiovascular Surgery
04 Faculty of Medicine > University Hospital Zurich > Institute of Anesthesiology
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:2014
Deposited On:04 Feb 2014 07:35
Last Modified:05 Apr 2016 17:30
Publisher:Europa Edition
ISSN:1774-024X
Free access at:Publisher DOI. An embargo period may apply.
Publisher DOI:https://doi.org/10.4244/EIJV9I10A203
Official URL:http://www.pcronline.com/eurointervention/ahead_of_print/201310-02/
PubMed ID:24103772

Download

Full text not available from this repository.
View at publisher

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.