Integrated circuit detector technology in abdominal CT: Added value in obese patients


Morsbach, Fabian; Bickelhaupt, Sebastian; Rätzer, Susan; Schmidt, Bernhard; Alkadhi, Hatem (2014). Integrated circuit detector technology in abdominal CT: Added value in obese patients. American Journal of Roentgenology, 202(2):368-374.

Abstract

OBJECTIVE. The purpose of this article was to assess the effect of an integrated circuit (IC) detector for abdominal CT on image quality. MATERIALS AND METHODS. In the first study part, an abdominal phantom was scanned with various extension rings using a CT scanner equipped with a conventional discrete circuit (DC) detector and on the same scanner with an IC detector (120 kVp, 150 effective mAs, and 75 effective mAs). In the second study part, 20 patients were included who underwent abdominal CT both with the IC detector and previously at similar protocol parameters (120 kVp tube current-time product and 150 reference mAs using automated tube current modulation) with the DC detector. Images were reconstructed with filtered back projection. RESULTS. Image quality in the phantom was higher for images acquired with the IC compared with the DC detector. There was a gradually increasing noise reduction with increasing phantom sizes, with the highest (37% in the largest phantom) at 75 effective mAs (p < 0.001). In patients, noise was overall significantly (p = 0.025) reduced by 6.4% using the IC detector. Similar to the phantom, there was a gradual increase in noise reduction to 7.9% in patients with a body mass index of 25 kg/m(2) or lower (p = 0.008). Significant correlation was found in patients between noise and abdominal diameter in DC detector images (r = 0.604, p = 0.005), whereas no such correlation was found for the IC detector (r = 0.427, p = 0.060). CONCLUSION. Use of an IC detector in abdominal CT improves image quality and reduces image noise, particularly in overweight and obese patients. This noise reduction has the potential for dose reduction in abdominal CT.

Abstract

OBJECTIVE. The purpose of this article was to assess the effect of an integrated circuit (IC) detector for abdominal CT on image quality. MATERIALS AND METHODS. In the first study part, an abdominal phantom was scanned with various extension rings using a CT scanner equipped with a conventional discrete circuit (DC) detector and on the same scanner with an IC detector (120 kVp, 150 effective mAs, and 75 effective mAs). In the second study part, 20 patients were included who underwent abdominal CT both with the IC detector and previously at similar protocol parameters (120 kVp tube current-time product and 150 reference mAs using automated tube current modulation) with the DC detector. Images were reconstructed with filtered back projection. RESULTS. Image quality in the phantom was higher for images acquired with the IC compared with the DC detector. There was a gradually increasing noise reduction with increasing phantom sizes, with the highest (37% in the largest phantom) at 75 effective mAs (p < 0.001). In patients, noise was overall significantly (p = 0.025) reduced by 6.4% using the IC detector. Similar to the phantom, there was a gradual increase in noise reduction to 7.9% in patients with a body mass index of 25 kg/m(2) or lower (p = 0.008). Significant correlation was found in patients between noise and abdominal diameter in DC detector images (r = 0.604, p = 0.005), whereas no such correlation was found for the IC detector (r = 0.427, p = 0.060). CONCLUSION. Use of an IC detector in abdominal CT improves image quality and reduces image noise, particularly in overweight and obese patients. This noise reduction has the potential for dose reduction in abdominal CT.

Citations

7 citations in Web of Science®
7 citations in Scopus®
Google Scholar™

Altmetrics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Diagnostic and Interventional Radiology
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:2014
Deposited On:28 Feb 2014 15:55
Last Modified:05 Apr 2016 17:31
Publisher:American Roentgen Ray Society
ISSN:0361-803X
Free access at:Publisher DOI. An embargo period may apply.
Publisher DOI:https://doi.org/10.2214/AJR.13.10839
PubMed ID:24450679

Download

Full text not available from this repository.
View at publisher

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.

Author Collaborations