Cardiac output measurement by pulse dye densitometry in cardiac surgery


Baulig, W; Bernhard, E O; Bettex, D; Schmidlin, D; Schmid, E R (2005). Cardiac output measurement by pulse dye densitometry in cardiac surgery. Anaesthesia, 60(10):968-973.

Abstract

Summary The aim of this study was to compare the accuracy of pulse dye densitometry with that of bolus thermodilution cardiac output measurement in patients before and after elective coronary artery bypass grafting. Twenty-eight patients were studied. Agreement between mean thermodilution and pulse dye densitometry cardiac output values was assessed by Bland-Altman analysis. Preoperative median [range] cardiac output was 3.87 [2.37-6.0] l.min(-1) by thermodilution, and 3.11 [1.7-5.45] l.min(-1) by pulse dye densitometry using indocyanine green 5 mg. Pulse dye densitometry underestimated cardiac output (mean bias - 0.42 l.min(-1)); the limits of agreement were +/- 1.91 l.min(-1), and mean error was 50.3%, indicating low precision. Preoperative median [range] cardiac output was 3.85 [2.2-6.0] l.min(-1) for bolus thermodilution cardiac output and 4.2 [2.0-7.2] l.min(-1) for pulse dye densitometry using indocyanine green 20 mg. Mean bias was + 0.566 l.min(-1), the limits of agreement were +/- 2.51 l.min(-1) and mean error was 60.9%. Postoperative cardiac output data were not analysed because pulse dye densitometry signals were low or absent in > 50% of the patients. We conclude that pulse dye densitometry using indocyanine green 5 mg or 20 mg is inaccurate in anaesthetised patients before coronary artery bypass surgery and cannot be used after surgery because of a high incidence of low pulse dye densitometry signal amplitudes.

Abstract

Summary The aim of this study was to compare the accuracy of pulse dye densitometry with that of bolus thermodilution cardiac output measurement in patients before and after elective coronary artery bypass grafting. Twenty-eight patients were studied. Agreement between mean thermodilution and pulse dye densitometry cardiac output values was assessed by Bland-Altman analysis. Preoperative median [range] cardiac output was 3.87 [2.37-6.0] l.min(-1) by thermodilution, and 3.11 [1.7-5.45] l.min(-1) by pulse dye densitometry using indocyanine green 5 mg. Pulse dye densitometry underestimated cardiac output (mean bias - 0.42 l.min(-1)); the limits of agreement were +/- 1.91 l.min(-1), and mean error was 50.3%, indicating low precision. Preoperative median [range] cardiac output was 3.85 [2.2-6.0] l.min(-1) for bolus thermodilution cardiac output and 4.2 [2.0-7.2] l.min(-1) for pulse dye densitometry using indocyanine green 20 mg. Mean bias was + 0.566 l.min(-1), the limits of agreement were +/- 2.51 l.min(-1) and mean error was 60.9%. Postoperative cardiac output data were not analysed because pulse dye densitometry signals were low or absent in > 50% of the patients. We conclude that pulse dye densitometry using indocyanine green 5 mg or 20 mg is inaccurate in anaesthetised patients before coronary artery bypass surgery and cannot be used after surgery because of a high incidence of low pulse dye densitometry signal amplitudes.

Citations

24 citations in Web of Science®
28 citations in Scopus®
Google Scholar™

Altmetrics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Institute of Anesthesiology
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:2005
Deposited On:18 Mar 2014 16:08
Last Modified:05 Apr 2016 17:38
Publisher:Wiley-Blackwell
ISSN:0003-2409
Free access at:Publisher DOI. An embargo period may apply.
Publisher DOI:https://doi.org/10.1111/j.1365-2044.2005.04296.x
PubMed ID:16179040

Download

Full text not available from this repository.
View at publisher

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.

Author Collaborations