UZH-Logo

Maintenance Infos

Medication adherence assessment: high accuracy of the new Ingestible Sensor System in kidney transplants


Eisenberger, Ute; Wüthrich, Rudolf P; Bock, Andreas; Ambühl, Patrice; Steiger, Jürg; Intondi, Allison; Kuranoff, Susan; Maier, Thomas; Green, Damian; DiCarlo, Lorenzo; Feutren, Gilles; De Geest, Sabina (2013). Medication adherence assessment: high accuracy of the new Ingestible Sensor System in kidney transplants. Transplantation, 96(3):245-250.

Abstract

BACKGROUND: This open-label single-arm exploratory study evaluated the accuracy of the Ingestible Sensor System (ISS), a novel technology for directly assessing the ingestion of oral medications and treatment adherence.
METHODS: ISS consists of an ingestible event marker (IEM), a microsensor that becomes activated in gastric fluid, and an adhesive personal monitor (APM) that detects IEM activation. In this study, the IEM was combined to enteric-coated mycophenolate sodium (ECMPS). Twenty stable adult kidney transplants received IEM-ECMPS for a mean of 9.2 weeks totaling 1227 cumulative days.
RESULTS: Eight patients prematurely discontinued treatment due to ECMPS gastrointestinal symptoms (n=2), skin intolerance to APM (n=2), and insufficient system usability (n=4). Rash or erythema due to APM was reported in 7 (37%) patients, all during the first month of use. No serious or severe adverse events and no rejection episode were reported. IEM detection accuracy was 100% over 34 directly observed ingestions; Taking Adherence was 99.4% over a total of 2824 prescribed IEM-ECMPS ingestions. ISS could detect accurately the ingestion of two IEM-ECMPS capsules taken at the same time (detection rate of 99.3%, n=2376).
CONCLUSIONS: ISS is a promising new technology that provides highly reliable measurements of intake and timing of intake of drugs that are combined with the IEM.

BACKGROUND: This open-label single-arm exploratory study evaluated the accuracy of the Ingestible Sensor System (ISS), a novel technology for directly assessing the ingestion of oral medications and treatment adherence.
METHODS: ISS consists of an ingestible event marker (IEM), a microsensor that becomes activated in gastric fluid, and an adhesive personal monitor (APM) that detects IEM activation. In this study, the IEM was combined to enteric-coated mycophenolate sodium (ECMPS). Twenty stable adult kidney transplants received IEM-ECMPS for a mean of 9.2 weeks totaling 1227 cumulative days.
RESULTS: Eight patients prematurely discontinued treatment due to ECMPS gastrointestinal symptoms (n=2), skin intolerance to APM (n=2), and insufficient system usability (n=4). Rash or erythema due to APM was reported in 7 (37%) patients, all during the first month of use. No serious or severe adverse events and no rejection episode were reported. IEM detection accuracy was 100% over 34 directly observed ingestions; Taking Adherence was 99.4% over a total of 2824 prescribed IEM-ECMPS ingestions. ISS could detect accurately the ingestion of two IEM-ECMPS capsules taken at the same time (detection rate of 99.3%, n=2376).
CONCLUSIONS: ISS is a promising new technology that provides highly reliable measurements of intake and timing of intake of drugs that are combined with the IEM.

Citations

12 citations in Web of Science®
15 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

75 downloads since deposited on 14 Feb 2014
33 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Nephrology
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:2013
Deposited On:14 Feb 2014 09:04
Last Modified:05 Apr 2016 17:40
Publisher:Lippincott, Williams & Wilkins
ISSN:0041-1337
Free access at:Publisher DOI. An embargo period may apply.
Publisher DOI:https://doi.org/10.1097/TP.0b013e31829b7571
PubMed ID:23823651
Permanent URL: https://doi.org/10.5167/uzh-92740

Download

[img]
Preview
Content: Published Version
Filetype: PDF
Size: 2MB
View at publisher

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.