de Vasconcellos, Diego Klee; Kojima, Alberto Noriyuki; Mesquita, Alfredo Mikail Melo; Bottino, Marco Antonio; Özcan, Mutlu (2014). A microstrain comparison of passively fitting screw-retained and cemented titanium frameworks. Journal of Prosthetic Dentistry, 112(4):834-838.

Abstract

STATEMENT OF PROBLEM An imprecise fit between frameworks and supporting dental implants in loaded protocols increases the strain transferred to the periimplant bone, which may impair healing or generate microgaps. PURPOSE The purpose of this study was to investigate the microstrain between premachined 1-piece screw-retained frameworks (group STF) and screw-retained frameworks fabricated by cementing titanium cylinders to the prefabricated framework (group CTF). This procedure was developed to correct the misfit between frameworks and loaded implants. MATERIAL AND METHODS Four internal hexagon cylindrical implants were placed 10 mm apart in a polyurethane block by using the surgical guides of the corresponding implant system. Previously fabricated titanium frameworks (n=10) were divided into 2 groups. In group STF, prefabricated machined frameworks were used (n=5), and, in group CTF, the frameworks were fabricated by using a passive fit procedure, which was developed to correct the misfit between the cast titanium frameworks and supporting dental implants (n=5). Both groups were screw-retained under torque control (10 Ncm). Six strain gauges were placed on the upper surface of the polyurethane block, and 3 strain measurements were recorded for each framework. Data were analyzed with the Student t test (α=.05). RESULTS The mean microstrain values between the framework and the implants were significantly higher for group STF (2517 mε) than for group CTF (844 mε) (P<.05). CONCLUSIONS Complete-arch implant frameworks designed for load application and fabricated by using the passive fit procedure decreased the strain between the frameworks and implants more than 1 piece prefabricated machined frameworks.

Abstract

STATEMENT OF PROBLEM An imprecise fit between frameworks and supporting dental implants in loaded protocols increases the strain transferred to the periimplant bone, which may impair healing or generate microgaps. PURPOSE The purpose of this study was to investigate the microstrain between premachined 1-piece screw-retained frameworks (group STF) and screw-retained frameworks fabricated by cementing titanium cylinders to the prefabricated framework (group CTF). This procedure was developed to correct the misfit between frameworks and loaded implants. MATERIAL AND METHODS Four internal hexagon cylindrical implants were placed 10 mm apart in a polyurethane block by using the surgical guides of the corresponding implant system. Previously fabricated titanium frameworks (n=10) were divided into 2 groups. In group STF, prefabricated machined frameworks were used (n=5), and, in group CTF, the frameworks were fabricated by using a passive fit procedure, which was developed to correct the misfit between the cast titanium frameworks and supporting dental implants (n=5). Both groups were screw-retained under torque control (10 Ncm). Six strain gauges were placed on the upper surface of the polyurethane block, and 3 strain measurements were recorded for each framework. Data were analyzed with the Student t test (α=.05). RESULTS The mean microstrain values between the framework and the implants were significantly higher for group STF (2517 mε) than for group CTF (844 mε) (P<.05). CONCLUSIONS Complete-arch implant frameworks designed for load application and fabricated by using the passive fit procedure decreased the strain between the frameworks and implants more than 1 piece prefabricated machined frameworks.

Statistics

Citations

3 citations in Web of Science®
3 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

39 downloads since deposited on 13 Nov 2014
15 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > Center for Dental Medicine > Clinic for Fixed and Removable Prosthodontics
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:October 2014
Deposited On:13 Nov 2014 12:37
Last Modified:24 Jan 2017 08:13
Publisher:Elsevier
ISSN:0022-3913
Publisher DOI:https://doi.org/10.1016/j.prosdent.2013.12.018
PubMed ID:24798096

Download

Preview Icon on Download
Content: Published Version
Filetype: PDF - Registered users only
Size: 1MB
View at publisher
Preview Icon on Download
Preview
Content: Accepted Version
Filetype: PDF
Size: 2MB

Article Networks

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.

Author Collaborations