Abstract

Fractures of the orbital wall and floor can be challenging due to the demanding three-dimensional anatomy and limited intraoperative overview. Misfitting implants and inaccurate surgical technique may lead to visual disturbance and unaesthetic results. A new approach using individually manufactured titanium implants (KLS Martin, Group, Germany) for daily routine is presented in the current paper. Preoperative CT-scan data were processed in iPlan 3.0.5 (Brainlab, Feldkirchen, Germany) to generate a 3D-reconstruction of the affected orbit using the mirrored non-affected orbit as template and the extent of the patient specific implant (PSI) was outlined and three landmarks were positioned on the planned implant in order to allow easy control of the implant's position by intraoperative navigation. Superimposition allows the comparison of the postoperative result with the preoperative planning. Neither reoperation was indicated due to malposition of the implant and the ocular bulb nor visual impairments could be assessed. PSI allows precise reconstruction of orbital fractures by using a complete digital workflow and should be considered superior to manually bent titanium mesh implants.

Abstract

Fractures of the orbital wall and floor can be challenging due to the demanding three-dimensional anatomy and limited intraoperative overview. Misfitting implants and inaccurate surgical technique may lead to visual disturbance and unaesthetic results. A new approach using individually manufactured titanium implants (KLS Martin, Group, Germany) for daily routine is presented in the current paper. Preoperative CT-scan data were processed in iPlan 3.0.5 (Brainlab, Feldkirchen, Germany) to generate a 3D-reconstruction of the affected orbit using the mirrored non-affected orbit as template and the extent of the patient specific implant (PSI) was outlined and three landmarks were positioned on the planned implant in order to allow easy control of the implant's position by intraoperative navigation. Superimposition allows the comparison of the postoperative result with the preoperative planning. Neither reoperation was indicated due to malposition of the implant and the ocular bulb nor visual impairments could be assessed. PSI allows precise reconstruction of orbital fractures by using a complete digital workflow and should be considered superior to manually bent titanium mesh implants.

Statistics

Citations

28 citations in Web of Science®
28 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

115 downloads since deposited on 30 Dec 2014
46 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > Center for Dental Medicine > Clinic for Cranio-Maxillofacial Surgery
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:2014
Deposited On:30 Dec 2014 13:17
Last Modified:08 Dec 2017 09:18
Publisher:Elsevier
ISSN:1010-5182
Publisher DOI:https://doi.org/10.1016/j.jcms.2014.10.024
PubMed ID:25465486

Download

Download PDF  'Patient specific implants (PSI) in reconstruction of orbital floor and wall fractures'.
Preview
Content: Accepted Version
Filetype: PDF
Size: 641kB
View at publisher