Toward more accurate measurements of anorectal motor and sensory function in routine clinical practice: Validation of High-Resolution Anorectal Manometry and Rapid Barostat Bag measurements of rectal function


Sauter, M; Heinrich, H; Fox, Mark; Misselwitz, B; Halama, M; Schwizer, W; Fried, M; Fruehauf, H (2014). Toward more accurate measurements of anorectal motor and sensory function in routine clinical practice: Validation of High-Resolution Anorectal Manometry and Rapid Barostat Bag measurements of rectal function. Neurogastroenterology and Motility, 26(5):685-695.

Abstract

BACKGROUND: Measurements of anorectal function using high-resolution anorectal manometry (HR-ARM) and rectal barostat technology provide more reliable results than standard ARM with an elastic balloon; however, HR-ARM results have not been compared to ARM and standard barostat protocols are impractical in routine clinical practice. The aim of this study was to validate HR-ARM against standard ARM and standard barostat against a novel Rapid Barostat Bag (RBB) measurement and elastic balloon measurements of rectal function.
METHODS: Twenty-six healthy volunteers (15 female, 11 male, 19-52 years) were studied. Measurements of anal function and simulated defecation were compared for 12-sensor HR-ARM and 6-sensor standard ARM using line plots from the same recording. Rectal capacity, compliance, and sensation (volume threshold) were measured by elastic balloon, standard barostat, and RBB methods using stepwise inflation of a 700-mL polyethylene bag to 40 mmHg distension by electronic barostat and handheld syringe monitored by sphygmo-manometer, respectively. Results are reported as mean ± SD. Bland-Altman plots and correlation coefficients (r) for measurements were calculated.
KEY RESULTS: There was excellent agreement between HR- and standard ARM measurements (r > 0.86, <25 mmHg difference) and between standard barostat and RBB measurements of rectal capacity (r = 0.97, <25 mL difference). Correlation coefficients of threshold volumes for initial perception, urgency and discomfort were 0.37, 0.71, and 0.95, respectively. No significant correlation was present with elastic balloon measurements. Time to complete studies was shorter for HR-ARM than standard ARM and for RBB than standard barostat in historical controls.
CONCLUSIONS AND INFERENCES: HR-ARM with RBB measurements of anorectal function provides quick and reasonably accurate measurements of continence function suitable for use in routine clinical practice (ClinicalTrial.gov NCT01456442).

Abstract

BACKGROUND: Measurements of anorectal function using high-resolution anorectal manometry (HR-ARM) and rectal barostat technology provide more reliable results than standard ARM with an elastic balloon; however, HR-ARM results have not been compared to ARM and standard barostat protocols are impractical in routine clinical practice. The aim of this study was to validate HR-ARM against standard ARM and standard barostat against a novel Rapid Barostat Bag (RBB) measurement and elastic balloon measurements of rectal function.
METHODS: Twenty-six healthy volunteers (15 female, 11 male, 19-52 years) were studied. Measurements of anal function and simulated defecation were compared for 12-sensor HR-ARM and 6-sensor standard ARM using line plots from the same recording. Rectal capacity, compliance, and sensation (volume threshold) were measured by elastic balloon, standard barostat, and RBB methods using stepwise inflation of a 700-mL polyethylene bag to 40 mmHg distension by electronic barostat and handheld syringe monitored by sphygmo-manometer, respectively. Results are reported as mean ± SD. Bland-Altman plots and correlation coefficients (r) for measurements were calculated.
KEY RESULTS: There was excellent agreement between HR- and standard ARM measurements (r > 0.86, <25 mmHg difference) and between standard barostat and RBB measurements of rectal capacity (r = 0.97, <25 mL difference). Correlation coefficients of threshold volumes for initial perception, urgency and discomfort were 0.37, 0.71, and 0.95, respectively. No significant correlation was present with elastic balloon measurements. Time to complete studies was shorter for HR-ARM than standard ARM and for RBB than standard barostat in historical controls.
CONCLUSIONS AND INFERENCES: HR-ARM with RBB measurements of anorectal function provides quick and reasonably accurate measurements of continence function suitable for use in routine clinical practice (ClinicalTrial.gov NCT01456442).

Statistics

Citations

13 citations in Web of Science®
14 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

0 downloads since deposited on 07 Jan 2015
0 downloads since 12 months

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Gastroenterology and Hepatology
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:2014
Deposited On:07 Jan 2015 08:44
Last Modified:05 Apr 2016 18:41
Publisher:Wiley-Blackwell Publishing, Inc.
ISSN:1350-1925
Publisher DOI:https://doi.org/10.1111/nmo.12317
PubMed ID:24517865

Download

Preview Icon on Download
Content: Published Version
Filetype: PDF - Registered users only
Size: 496kB
View at publisher

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.

Author Collaborations