Model-based iterative reconstruction for improvement of low-contrast detectability in liver CT at reduced radiation dose: Ex-vivo experience


Husarik, D B; Alkadhi, H; Puippe, G D; Reiner, C S; Chuck, N C; Morsbach, F; Szucs-Farkas, Z; Schindera, S T (2015). Model-based iterative reconstruction for improvement of low-contrast detectability in liver CT at reduced radiation dose: Ex-vivo experience. Clinical Radiology, 70(4):366-372.

Abstract

AIM: To compare low-contrast detectability, and qualitative and quantitative image parameters on standard and reduced radiation dose abdominal CT reconstructed with filtered back projection (FBP) and model-based iterative reconstruction (MBIR).
MATERIALS AND METHODS: A custom built liver phantom containing 43 lesions was imaged at 120 kVp and four radiation dose levels (100% = 188 mAs, 50%, 25%, and 10%). Image noise and contrast-to-noise ratios (CNR) were assessed. Lesion detection and qualitative image analysis (five-point Likert scale with 1 = worst, 5 = best for confidence) was performed by three independent radiologists.
RESULTS: CNR on MBIR images was significantly higher (mean 246%, range 151-383%) and image noise was significantly lower (69%, 59-78%) than on FBP images at the same radiation dose (both p < 0.05). On MBIR 10% images, CNR (3.3 ± 0.3) was significantly higher and noise (15 ± 1HU) significantly lower than on FBP 100% images (2.5 ± 0.1; 21 ± 1 HU). On 100% images, lesion attenuation was significantly lower with MBIR than with FBP (mean difference -2 HU). Low-contrast detectability and qualitative results were similar with MBIR 50% and FBP 100%.
CONCLUSION: Low-contrast detectability with MBIR 50% and FBP 100% were equal. Quantitative parameters on even lower dose MBIR images are superior to 100%-dose FBP images. Some attenuation values differ significantly with MBIR compared with FBP.

Abstract

AIM: To compare low-contrast detectability, and qualitative and quantitative image parameters on standard and reduced radiation dose abdominal CT reconstructed with filtered back projection (FBP) and model-based iterative reconstruction (MBIR).
MATERIALS AND METHODS: A custom built liver phantom containing 43 lesions was imaged at 120 kVp and four radiation dose levels (100% = 188 mAs, 50%, 25%, and 10%). Image noise and contrast-to-noise ratios (CNR) were assessed. Lesion detection and qualitative image analysis (five-point Likert scale with 1 = worst, 5 = best for confidence) was performed by three independent radiologists.
RESULTS: CNR on MBIR images was significantly higher (mean 246%, range 151-383%) and image noise was significantly lower (69%, 59-78%) than on FBP images at the same radiation dose (both p < 0.05). On MBIR 10% images, CNR (3.3 ± 0.3) was significantly higher and noise (15 ± 1HU) significantly lower than on FBP 100% images (2.5 ± 0.1; 21 ± 1 HU). On 100% images, lesion attenuation was significantly lower with MBIR than with FBP (mean difference -2 HU). Low-contrast detectability and qualitative results were similar with MBIR 50% and FBP 100%.
CONCLUSION: Low-contrast detectability with MBIR 50% and FBP 100% were equal. Quantitative parameters on even lower dose MBIR images are superior to 100%-dose FBP images. Some attenuation values differ significantly with MBIR compared with FBP.

Statistics

Citations

5 citations in Web of Science®
5 citations in Scopus®
Google Scholar™

Altmetrics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Diagnostic and Interventional Radiology
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:2015
Deposited On:15 Jan 2015 11:37
Last Modified:05 Apr 2016 18:44
Publisher:Elsevier
ISSN:0009-9260
Publisher DOI:https://doi.org/10.1016/j.crad.2014.11.015
PubMed ID:25554541

Download

Full text not available from this repository.
View at publisher

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.